Specification Information
Transportation and
storage conditions
▪
Ambient temperature range of -40°F to +159°F (-40°C to +70°C).
▪
Relative humidity range of 10% to 95%, non-condensing.
▪
Atmospheric pressure range of 7.25 psi to 15.37 psi (50kPa to 106kPa).
Operating conditions
▪
The warmer must acclimate to the room temperature it will be placed in—24
hours is recommended.
▪
The recommended environmental temperature range is 60°F to 90°F
(15°C to 32°C).
▪
The recommended relative humidity is above 20%, non-condensing.
Standards for
electrical equipment
▪
Medical equipment listed by Underwriters Laboratories with respect to
electrical shock, fire, and mechanical hazards only, in accordance with UL
61010-1 and CAN/CSA C22.2 No. 61010-1.
▪
Grounding reliability can only be achieved when the appliance is
connected to an equivalent receptacle marked “Hospital Grade.”
▪
EN61326-1; EN61000-3-2; EN61000-3-3
▪
IP-XO ordinary
Model
EC1350BL
EC1850BL
Intended use
Bottom cavity—warming blankets
TriMode cavity—warming blankets, irrigation or injection fluids
Weight
Net: 336 lb (152 kg)
Ship: 407 lb (185 kg)
Net: 457 lb (207 kg)
Ship: 528 lb (239 kg)
Storage cavity
capacity
TriMode cavity: 24 1-liter
bottles or 20 1-liter bags, or
9–12 blankets (3.9 ft
3
)
Blanket cavity: 24–30 blankets
(9.4 ft
3
)
TriMode cavity: 30 1-liter
bottles or 24 1-liter bags, or
10–14 blankets (5.4 ft
3
)
Blanket cavity: 30–40 blankets
(13 ft
3
)
Temperature
range
Irrigation fluids (IRR):
90°F to 150°F (32°C to 66°C)
Injection fluids (INJ):
90°F to 104°F (32°C to 40°C)
Blanket: 90°F to 160°F (32°C to 71°C)
Clearance
requirements
2” (102mm) from rear
1” (25mm) from top
1” (51mm) from sides
NOTE:
See topic
Dimension
Drawings
for EC1350BL Built-
In requirements.
4” (102mm) from rear
1” (25mm) from top
2” (51mm) from sides
1/2” (13mm) from bottom
E471516
Protective Earth
Ground Symbol