18
OPTIMIZER ST2 FILTER MANIFOLD
Installation and Use Manual | Entegris, Inc.
ATTACHING THE FILTER OR FLUSHING SHELL
—
1. Raise the manifold lever completely to the top.
2. Place the filter or flushing shell in the groove of
the filter guide and push it in along the guide until
you hear it click into place. (Though there may be
some resistance, push the filter or flushing shell
to the end.)
CAUTION: If the lever is not raised enough, the
filter or flushing shell cannot be inserted along
the filter guide.
Remove the filter or flushing shell, lift the lever
to the top, and install again.
If the filter or flushing shell is forcibly inserted
while the lever is not completely raised, the
manifold or filter may be damaged or chemical
may leak.
3. Lower the lever to position 2. If you feel resistance
stop lowering the level and confirm the filter is
inserted completely.