50455/RTEX/0807
UTImeter Rtex
40
9.4.3 Disconnecting the tape from the reel
axle
Remove the external reel flange (3 screws to
uns crew with the 2.5 Allen key).
Remove the axle cover (3 screws to unscrew
with the 2.5 Allen key).
Unscrew with the 2.5 Allen key the 4 screws
(K) of the washer holder, as shown on Figure
9-7.
Remove the tape from the reel axle.
9.4.4 Removing the tape from the frame
Remove the tape protection tube from the tape.
Turn the tape cleaner in position "DOWN" to
free the tape.
Pull the tape gently out of the tape cleaner.
Pull the tape adaptor end out of the housing,
through the storage tube.
Slacken the tape a few turns from the reel axle.
Remove the tape from the housing.
9.4.5 Mounting the new tape
Install the new tape on the reel axle.
Leave approximatively 20 cm of tape free at
the core.
Make a loop (M) and a S-shape (L) with the
tape as shown on Figure 9-7.
Pass the tape end through the axle core.
Secure the gaskets and the washers mounted
on the tape in the axle core with the washer
holder and its 4 screws (K) as shown on Figure
9-7.
On the electronic box side, adjust the black
oversleeve just to the edge of the tape holder
(pull the tape gently from the other side) and
tighten the tape end as shown on Figure 9-6
with.
Follow in the reverse order the instructions of
sub-section 9.4.2 to re-install the electronic
box.
If necessary, readjust the loop (M) and the S-
shape (L) of the tape at the core of the reel
axle.
Follow the instructions of sub-section 9.4.4 in
the reverse order to pass the tape through the
tape cleaner, the storage tube and to mount
the tape protection tube on.
Put back the reel flange and its 3 securing
screws.
Follow the instructions of section 9.5 "Sensing
probe replacement" to re-install the sensor on
the tape.
Carry out the functional tests as per 8.8
"Checking the functions before using the
instrument".
If there is any problem, refer to section 10
"Trouble shooting".
(K) securing screws (4 x)
(L) s-shaped stopping device
(M) tape loop
Figure 9-7
Summary of Contents for HERMetic UTImeter Rtex
Page 55: ...50455 RTEX 0807 UTImeter Rtex 55 Figure 12 1 general assembly list of the main spare parts...
Page 56: ...50455 RTEX 0807 UTImeter Rtex 56 Figure 12 2 display unit assembly TS 10210 details...
Page 59: ...50455 RTEX 0807 UTImeter Rtex 59 Figure 12 5 storage tube SS1 Q1 TS 10184 details...
Page 60: ...50455 RTEX 0807 UTImeter Rtex 60 Figure 12 6 storage tube SS1 Q2 TS 10183 details...
Page 61: ...50455 RTEX 0807 UTImeter Rtex 61 Figure 12 7 storage tube SS2 Q2 with load TS 10182 details...
Page 62: ...50455 RTEX 0807 UTImeter Rtex 62 Figure 12 8 tape cleaner details...
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......