8.9 Installation of the instrument
This HERMetic equipment must be coupled to
a certified HERMetic valve.
Before
starting
please
read
c arefully the
chapter “Recommendation for safe use” and
follow your company's safety instructions.
Check that the HERMetic valve is closed.
Remove the end cap (weather cap / blind cover
/ security cover) of the HERMetic valve.
Clean the seal surfaces of the nipple of the
valve and of the coupler of the instrument from
dust or grease.
Note:
Cleaning
of the mating surfaces is very
important for
earth grounding
purpose
and
for
good
accuracy
on
zero
reference level
.
Check whether the tape protection tube is
moving freely.
Install the HERMetic instrument on top of the
valve by means of the quick coupler. Ensure
that the equipment is properly earthed. If not,
ground it with the (optional) grounding cable
before operating.
8.10 Ullage / interface
measurement
Install the HERMetic equipment as per 8.9
"Installation of the instrument”.
Open the valve by turning the handle.
Switch on the equipment: a control beep is
audible every 2 seconds.
Put the tape cleaner on the "DOWN" position.
Disengage the knob of the crank handle and
lower the sensing probe into the tank by turning
the reel. Make sure that the tape does not rub
on any sharp edge when lowering as its
insulation could be damaged.
As soon as the sensor comes in contact with
the petroleum product the control beep will
change for a continuous beep. Raise the
sensing probe again until the continuous beep
stops and lower the sensing probe again slowly
until the continuous beep is heard again. Now
the ullage level can be read against the ullage
reference. If the zero-ullage reference does not
correspond
to
the
reading
index
of
the
instrument, a correction has to be made
accordingly.
Lower the sensing probe further until the
sensor touches the oil-water interface. As soon
as the sensor comes in contact with water the
continuous beep will change for an intermittent
beep. The difference between the ullage
reading and the interface reading represents
the thickness of the product layer.
When
the
measurements
are
completed,
switch off the unit, turn the tape cleaner on
"UP" position and wind up the tape until the
sensing probe is in the storage tube. The
reading on the tape shall be less than 420 mm
or 1 ft 5 inch.
Close the valve and disconnect the instrument
from the nipple.
Put the end cap back on the valve.
Summary of Contents for HERMetic UTImeter Rtex
Page 55: ...50455 RTEX 0807 UTImeter Rtex 55 Figure 12 1 general assembly list of the main spare parts...
Page 56: ...50455 RTEX 0807 UTImeter Rtex 56 Figure 12 2 display unit assembly TS 10210 details...
Page 59: ...50455 RTEX 0807 UTImeter Rtex 59 Figure 12 5 storage tube SS1 Q1 TS 10184 details...
Page 60: ...50455 RTEX 0807 UTImeter Rtex 60 Figure 12 6 storage tube SS1 Q2 TS 10183 details...
Page 61: ...50455 RTEX 0807 UTImeter Rtex 61 Figure 12 7 storage tube SS2 Q2 with load TS 10182 details...
Page 62: ...50455 RTEX 0807 UTImeter Rtex 62 Figure 12 8 tape cleaner details...
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......