Enraf-Nonius B.V.
Radarmed 650 Service Manual
1409773
Introduction
January 2000
1-4
The intensity can be adjusted step-less from about 5 to 150 watts and is shown on the LED
output power at the front panel meter to assist the operator.
The Radarmed 650 is suitable for use with the circular-field, through emitter and
longitudinal-field emitters.
Certain minimum distances from the surface of the body must be maintained with different
emitters.
- Circular-field emitter
approx. 10 cm
- Longitudinal-field emitter
approx. 5 cm
- Through emitter
0 cm
The required distance for the through emitter is determined by the internal design.
Circular-field emitter
The circular-field emitter can be used on restricted body areas to effect deeper penetration. It
is the most universal, and therefore the most used, emitter.
Longitudinal-field emitter
The longitudinal-field emitter is used to treat long parts of the body, such as limbs.
Through emitter
The through emitter is used to treat large areas. The special form of this emitter is well
adapted to the appropriate body contours. Restricted regions, such as the shoulder and knee
joints, can likewise be advantageously treated with this emitter, as can large muscular areas
such as the pectoral girdle. For reasons of hygiene, a cloth should always be placed
between the patient and emitter.
The Radarmed 650 has been designed and manufactured to ensure the highest level of
safety; the unit fully complies with the IEC 601-1 (general standard), IEC 601-2-6 (microwave
equipment) and IEC 601-1-2 (EMC) standards.
1.3
TECHNICAL DATA
1.3.1
General
Mains supply
: 230 VAC
Permissible mains voltage var.
: ±10%
Mains frequency
: 50-60 Hz
Current consumption
: max. 3.5 A at 230 V
Mains fuses
: 2 x T 6.3 H 250 V
Internal fuses
: T 0.315 A
(F103)
: T 0.5 A
(F104)
: T 0.2 A
(F105)
: T 1 A
(F106)
Patient leakage current
: typical < 10
A
(IEC requirement
100
A*)
Ditto, single fault condition
: typical < 10
A
(IEC requirement
500
A*)
Earth leakage current
: typical 250
A
(IEC requirement
500
A*)
Ditto, single fault condition
: typical 500
A
(IEC requirement
1000
A*)
Earth resistance
: 0.06
6
(typical)
(IEC601 requirement
0.2
6
)
Summary of Contents for Radarmed 650
Page 1: ...Radarmed 650 SERVICE MANUAL Part number 1409773...
Page 2: ......
Page 6: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Index January 2000 4...
Page 12: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Introduction January 2000 1 6...
Page 16: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Description January 2000 2 4...
Page 18: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Test equipment and tools January 2000 3 2...
Page 26: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Maintenance January 2000 4 8...
Page 28: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Storage and transport January 2000 5 2...
Page 38: ...Enraf Nonius B V Radarmed 650 Service Manual 1409773 Appendix B January 2000 B 6...
Page 51: ......
Page 52: ...Subsidiary of Henley Healthcare Inc Enraf Nonius B V P O Box 810 2600 AV Delft The Netherlands...