background image

 

Page 6 of 14 

EN109-1560750-44

 

6

 

Precautionary instructions  

  If the use of this device may have caused or contributed to an undesirable event such as 

death or serious injury to the user, the manufacturer AND the competent authority of the 
Member State MUST be notified immediately! 

  Read this Instructions for Use carefully before using the equipment. 
  Do not lift the couch by the table top parts or the all-around control. 

  Immediately after unpacking the couch, check whether the package contains all the items 

listed above, and check the couch for any damage. 
In case of damage, contact your local representative. Do NOT use the equipment! 

  Avoid exposure to direct sunlight, UV radiation, rain, excessive dust, moisture, mechanical 

vibrations, and shocks. 

  The Manumed ST couches are not intended for use in so-

called “wet rooms” (hydrotherapy 

rooms). 

  The device may not be used in close proximity (i.e. less than 2 metres) to shortwave or device. 

  The Manumed ST should be placed on a flat surface. 

  Turn the adjustable foot until it touches the ground and the couch is standing stable. 
  For correct assembly and safe use of possible optional accessories one must consult the 

separately provided documents included with the optional accessories. 

  Mains supply connections must comply with the national requirements regarding medical 

rooms. 

  Prior to connection of this device to the mains supply, check that the voltage and frequency 

stated on the type plate correspond with the available mains supply. 

  The Manumed ST is only meant to be used by competent personnel in physiotherapy, 

rehabilitation or adjacent areas. Avoid any use of the equipment by unauthorized personnel. 

  When used by children, supervision by a knowledgeable (para)medic is obligatory. At all times 

keep children away from the equipment when no supervision is at hand or use a safety switch. 

  Use both hands to change the couch settings: one hand for unlocking and the other to adjust 

the section. 

 

Never adjust a section when it supports the full weight of the patient!

 Each section of 

this Manumed ST is supported by gas springs. As well as easing operation, they offer full 
support of the sections. However, these gas springs are not strong enough to lift a section 
while it supports the full weight of the patient. 

 

A maximum patient’s weight of 

150 kg

 is allowed when the height of the couch is being 

adjusted. 

  Be aware of the duty cycle of the motor. The motor is not designed for continuous operation. 

For every 2 minutes the motor is running, it needs 18 minutes to cool down. If you do use the 
motor considerably longer without allowing it to cool down sufficiently, a thermal cut shuts 
off the power and protects the motor against overheating. In that case, let the motor cool 
down until it functions again. 

  The couch is not anti-static. 
  For safety reasons, the wheels of the unit should be retracted before the beginning of a 

treatment. 

  Never let your full weight rest on the switch bar! 
  Ensure that you will never rest your feet on the all-around-operation. 

Especially when seated next to the couch it can lead to a dangerous situation in which your 
leg can be jammed between the section and the all-around-operation. 

  Do not sit on the head and leg section of the couch. 

Summary of Contents for Manumed ST

Page 1: ...Manumed ST EN109 1560750 44 APRIL 4 2022 Instructions for use...

Page 2: ...3 2 Symbols 3 3 Package contents 4 4 Installation 5 5 Intended use and intended user 5 6 Precautionary instructions 6 7 Operation 7 8 Safety instructions 9 9 Cleaning maintenance and troubleshooting...

Page 3: ...o Manumed ST 2 section head section Manumed ST 2 section back section Manumed ST 3 section Manumed ST accessories 2 Symbols Symbol used Description Follow the instructions in the Instructions for Use...

Page 4: ...d water by mechanical casings and electrical enclosures Double Insulation Type B applied part complying with the specified requirements to provide protection against electric shock particularly regard...

Page 5: ...d ST should be placed on a flat surface Turn the adjustable foot until it touches the ground and the couch is standing stable For correct assembly and safe use of possible optional accessories one mus...

Page 6: ...nt to be used by competent personnel in physiotherapy rehabilitation or adjacent areas Avoid any use of the equipment by unauthorized personnel When used by children supervision by a knowledgeable par...

Page 7: ...upports the full weight of the patient Each section of this Manumed ST is supported by gas springs As well as easing operation they offer full support of the sections However these gas springs are not...

Page 8: ...ed underneath the framework with your foot the required height of the couch can be set At the head section of the couch the height can be adjusted by turning the handle to the left or to the right Nev...

Page 9: ...eside the couch is set in flexion position combined with the lowest possible position Underneath the head section between the head section and floor Danger of tilting Do not sit on the head and leg se...

Page 10: ...clean water Disinfection Clean the couch as described above before proceeding to disinfect the couch Follow the disinfectant s instructions for use Never use a mixture of disinfectants or a disinfecta...

Page 11: ...service life This device will remain suitable for its intended use as long as it is subjected to yearly maintenance by a qualified service engineer as described in the service manual and the service...

Page 12: ...EN ISO 10993 5 DIN EN ISO 10993 10 Technical modifications reserved Safety and performance standards This equipment complies with all requirements of the Medical Device Regulation EU 2017 745 Medical...

Page 13: ...rs be liable for any indirect special incidental or consequential damages arising from the use of or inability to use the product including without limitation damages for loss of goodwill work and pro...

Page 14: ...EN109 1560750 44 Copyright Enraf Nonius B V Vareseweg 127 3047 AT Rotterdam The Netherlands Tel 31 0 10 20 30 600 info enraf nonius nl...