53
The 4-series device is intended for operation in an electromagnetic environment where RF disturbances
are monitored. The customer or user of the 4-series device can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF telecommunications 4-
series (transmitters) and the 4-series device – according to the output power of the communications
device, as indicated below.
Rated output of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d= 1.2
√
P
80 MHz to 800 MHz
d= 0.35
√
P
800 MHz to 2.5 GHz
d= 0.7
√
P
0.01 0.12
0.035
0.07
0.1 0.38
0.11
0.22
1 1.2
0.35
0.70
10 3.8
1.1
2.2
100 12
3.5
7
For transmitters rated at a maximum output which is not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the respective column, whereby P is the maximum
rated output of the transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz the higher frequency range is applicable.
NOTE 2 These guidelines may not be applicable in all cases. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
11 Ordering information
Model numbers:
1498901
Sonopuls 490 with large ultrasound applicator
1498902
Sonopuls 490 with small ultrasound applicator
1498903
Sonopuls 490 with large and small ultrasound applicator
1498911
Sonopuls 492 with large ultrasound applicator
1498912
Sonopuls 492 with small ultrasound applicator
1498913
Sonopuls 492 with large and small ultrasound applicator
1498920 Endomed
482
1498950 Vacotron
460
Standard accessories Sonopuls 492:
1498757
4-Series Operating Instructions (cd rom)
1498756
4-Series Information Booklet
3444290 Mains
cable
230V-EUR
1498010
Device base (inclination support) (not for Vacotron 460)
xxxxxxx
Ultrasound multi-frequency treatment head(s) (model dependent)
xxxxxxx
Holder treatment head (model dependent)
3442929
(1)
Contact-gel, bottle 250 ml, 1x
3444021 Strap
250x3
cm
3444020 Strap
100x3
cm
2 x 3444129
Rubber electrodes 6x8 cm, 2 mm female, set of 2 pcs.
1460266
Moist pads for rubber electrodes 6x8 cm, set of 4 pcs.
2 x 3444211
Patient cable 2-core & 2 mm male plugs - black, with coloured clips
(1)
= The Sonopuls is delivered with 1 bottle of contact-gel. The article-
number 3442929 however represents a box of 12 bottles.