Water Reservoir Full
The water separation tank of the Vacotron is full. Continue the treatment with the standard electrodes or
empty the tank as follows:
•
Set power line switch [1] Off (0).
•
Detach the hose from the upper hose nipple [15] and empty the reservoir.
•
Reattach the hose to the hose nipple.
•
Set power line switch [1] On (1).
9.2.5 Vacuum
Leak
There probably is a leak in the vacuum system. This error is usually preceded by a continuously running
pump trying to reach the set vacuum. To protect the system, the pump is automatically stopped after a
certain amount of time. Inspect the vacuum cables and electrodes, set the vacuum back to zero and try
again. If the failure persists, stop using the device and contact your supplier.
9.2.6 Ultrasound
Applicator
Error
The ultrasound applicator has reported an error. Disconnect the device, wait some time and reconnect it.
If the error persists, stop using the device and contact your supplier.
9.2.7 Insufficient
DC
Supply
This problem can sometimes occur with the small ultrasound applicator when operating from the battery.
If possible, continue the treatment with reduced therapy levels or connect the apparatus to the mains
supply.
9.3 User
Maintenance
9.3.1
Optimize Contact Control Ultrasound Applicator
When you experience difficulties with the contact control function of the ultrasound applicator you can try
to resolve the problem as follows:
•
Ensure that the surface of the ultrasound applicator is clean and dry.
•
Place the ultrasound applicator in the holder.
•
Go to System Settings -> Maintenance and select Optimize Applicator A or B.
•
Touch the OK-button when the operation is complete.
9.3.2
Back-up and Restore Favorites
When you have programmed and stored several Favorites, you might want to make a back-up on an
external storage device. To store your favorites, proceed as follows:
•
Attach a USB-stick to the remote control connection [3]. Read and obey the warnings and
cautions mentioned in paragraph 8.1.
•
Go to Systems Settings -> Maintenance and select Back-up Favorites.
•
If an error occurs during the back-up operation, i.e. USB-stick full, this will be displayed in a pop-
up message.
•
Touch the OK-button when the operation is complete.
•
Detach the USB-stick
To restore your favorites:
•
Attach the USB-stick containing your Favorites to the remote control connection [3]. Read and
obey the warnings and cautions mentioned in paragraph 8.1.
•
Go to Systems Settings -> Maintenance and select Restore Favorites.
•
If an error occurs during the restore operation, i.e. no favorites found, this will be displayed in a
pop-up message.
•
Touch the OK-button when the operation is complete.