Page 18 of 46
EN109-1633750-40 IFU
Laser treatments must be delivered under the strict control of the operator to conscious
patients who are able to interact with the operator in response to stresses transmitted by the
machine. In case of non-compliance with the information provided, Enraf-Nonius B.V. will not
be held responsible for any accidents.
It is the responsibility of the operator to verify that the emitting head remains well adherent
to the treatment area, in order to avoid the emission of the laser in areas other than those to
be treated.
Never turn the handpiece to the parts of the body sensitive to the laser rays, for example the
eyes.
Do not look at the target beam area emitted by the handpiece or the direct reflections during
treatment without protective goggles.
The laser radiation coming out of the device is dangerous: always use the appropriate
protective goggles. Always avoid eye exposure to direct or diffuse laser radiation.
Before performing any user maintenance, switch off the device and disconnect the plug from
the Mains supply.
Do not spray the cleaning agent directly on the glass panel.
Do not use cleaning agents that contain strong alkalis, lye, acid, detergents with fluoride or
detergents with ammonia.
Use no liquid detergents, these can damage the device.
Do not allow any liquids to penetrate the device or its accessories while cleaning and
disinfecting!
Dry all sockets and connectors that have become wet before any further use!
Solvents used for cleaning and disinfection should be allowed to evaporate before the laser
equipment is used.
Mechanical shocks /or vibrations must be avoided while cleaning the device.
Do not drop the device.
The device should not be used when mechanical damage is noticed. This includes damaged
cable insulation.
In the event that an error message or warning appears, immediately stop all use of the device.
Shutdown and restart the device. If the problem persists contact the supplier or Enraf-Nonius
B.V. for service. Errors and warnings indicate an internal problem with the device that must be
tested by a field service technician certified by Enraf-Nonius B.V before any further operation
or use of the system.
Use of a device that indicates an error or warning may pose a risk of injury to the patient,
user, or extensive internal damage to the system.
In case of failure of the display or other obvious defects, unplug the device immediately and
notify a certified service technician.
In the case of ingress of liquids, unplug the device from the Mains supply and have it checked
by an authorized person.
Only service Technicians authorized by Enraf-Nonius B.V. can perform maintenance or modify
this equipment according to current Service Information.
If this equipment is modified, appropriate inspection and testing must be conducted to
ensure continued safe use of the equipment.
Maintenance shall not be performed on the equipment when equipment is in use or
connected to a patient.
Medical electrical devices such as the Endolaser 120 are subject to special precautions with
regard to electromagnetic compatibility (EMC) and must be installed and commissioned in
accordance with the EMC advice given in the Instructions For Use.
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are operating normally.