background image

GSM-60

 

ENMET 

 

Manual Revision Date – June 15, 2017

 

P a g e

 | 

3

 

Manual Part No. – 80003-600 

Rev. 1 – December 15, 2016 

 

1.0 Introduction 

The 

GSM-60

 

is a sample draw monitoring instrument that measures and detects certain toxic hazards gases.  The 

GSM-60

 was designed primarily 

for monitoring one area with a single sampling hose, either for an individual gas or multiple gases.  The instrument has one sampling pump and one 
flow sensor.  The instrument is available with up to 4 internal sensors.  Available sensors include, but are not, limited to carbon monoxide (CO), 
carbon dioxide (CO

2

), Hydrocarbons (HC) and variations in the oxygen (O

2

) content.  The sensors can be used alone or up to four sensors can be 

used together.  Some applications require monitoring of more than one area.  If two sampling hoses are connected to one 

GSM-60

 the flow sensor 

can only detect a total flow fault condition.  For example, a pump failure or both sampling hoses being blocked.  The system cannot detect a flow 
fault if only one of the two sampling hoses is blocked.   
 
Flowmeters should be installed in both sampling lines to provide a visual indication of proper flow.  In the instrument, a 24 V

DC

 sampling pump pass 

air over each sensor and the resulting electrical outputs are used to evaluate the air for the target gases. 
 
The 

GSM-60

 is a highly adaptable instrument.  Some adaptations require an addendum be added to the manual to facilitate use of the instruments 

with these adaptations.  If addendum is needed, see page(s) between sections replacement part numbers and warranty. 
 
Some features of the instruments are as follows: 

 

continuous monitoring of the sample air 

 

alarm relay contacts available on terminals 

 

continuous LCD display of gas and vapor concentrations 

 

a fault relay and visual fault alarm 

 

menu driven operational and maintenance controls 

 

low air flow fault indication and display 

 

menu driven calibration procedure 

 

alarm acknowledgement capability including audio defeat 

 

audio and visual alarms indicate unsafe conditions 

 

mA outputs for each target gas 

Hydrocarbons (HC) are limited to gases with an ionization potential of 10.6 eV or less.  See Appendix B for a list of gases and IPs. 

 
N

OTE

:

 

All specifications stated in this manual may change without notice.

 

 

1.1 Unpack 

Unpack the GSM-60 and examine it for shipping damage.  If such damage is observed, notify both ENMET customer service personnel and the 
commercial carrier involved immediately.

 

 

Regarding Damaged Shipments 

 

N

OTE

It is your responsibility to follow these instructions.  If they are not followed, the carrier will not honor any claims for damage.

 

 

 

This shipment was carefully inspected, verified and properly packaged at 

ENMET

 and delivered to the carrier in good condition. 

 

When it was picked up by the carrier at 

ENMET

, it legally became your company’s property. 

 

If your shipment arrives damaged: 

o

 

Keep the items, packing material, and carton “As Is.”  Within 5 days of receipt, notify the carrier’s local office and request immediate 
inspection of the carton and the contents. 

o

 

After the inspection and after you have received written acknowledgment of the damage from the carrier, contact 

ENMET

 Customer 

Service for return authorization and further instructions.  Please have your Purchase Order and Sales Order numbers available. 

 

ENMET

 either repairs or replaces damaged equipment and invoices the carrier to the extent of the liability coverage, usually $100.00.  

Repair or replacement charges above that value are your company’s responsibility. 

 

The shipping company may offer optional insurance coverage. 

ENMET

 only insures shipments with the shipping company when asked to do 

so in writing by our customer.  If you need your shipments insured, please forward a written request to 

ENMET

 Customer Service. 

 

Regarding Shortages 

If there are any shortages or questions regarding this shipment, please notify 

ENMET

 Customer Service within 5 days of receipt at the following 

address: 

ENMET 

680 Fairfield Court 

Ann Arbor, MI 48108 

734-761-1270 Fax 734-761-3220 

Toll Free: 800-521-2978 

1.2 Check Order 

Check, the contents of the shipment against the purchase order.  Verify that the GSM-60 is received as ordered.  If there are accessories on the 
order, ascertain that they are present.  Check the contents of calibration kits.  Notify ENMET customer service personnel of any discrepancy 
immediately. 

 

1.3 Serial Numbers  

Each GSM-60 is serialized.  These numbers are on tags on the equipment and are on record in an ENMET database. 

 

 

Summary of Contents for GSM-60

Page 1: ...ENMET 680 Fairfield Court Ann Arbor MI 48108 734 761 1270 Fax 734 761 3220 www enmet com GSM 60 Operation Manual...

Page 2: ...8 Hydrocarbon Sensor Response 14 5 0 MAINTENANCE 15 5 1 Cleaning Instructions 15 5 2 Maintenance Menu 15 5 2 1 Accessing Maintenance Menu 15 5 2 2 Maintenance Menu Flow Chart 16 5 3 Calibration for C...

Page 3: ...Figure 7 GSM 60 Calibration Connections 18 Figure 9 Location of Gas Sensor and PiD HC Manifolds 26 Figure 7 GSM 60 Calibration Connections 28 Figure 10 Carbon Monoxide Concentration 31 List of Tables...

Page 4: ...ppendix B for a list of gases and IPs NOTE All specifications stated in this manual may change without notice 1 1 Unpack Unpack the GSM 60 and examine it for shipping damage If such damage is observed...

Page 5: ...ample and calibration gas Luer fitting 2 2 Display Panel Features The display panel shown in Figure 1 is viewed through the clear front cover of the enclosure and is accessed by opening the cover Feat...

Page 6: ...Fitting Input Port Sample Calibration Audio Alarm Fault LED SELECT Switch Power LED OPTION Switch ALARM ACKNOWLEDGE ALARM DEFEAT Switch Visual Alarm O2 If Oxygen option is installed See note below Vi...

Page 7: ...e 2 Feature Description Relay Terminals This group of terminals is located at the left side of the Circuit Board For the contacts for each of four alarm relays and for the contacts of a fault relay Ou...

Page 8: ...k disconnect fittings are supplied for use with 1 8 ID tubing Mount the instrument on an appropriate vertical surface using the mounting flanges provided Avoid areas with excessive vibration or temper...

Page 9: ...are revision and gases monitored may be shown on the display The instrument may go into alarm briefly but the sensors stabilize quickly If the instrument persists in alarm acknowledge the alarm by pre...

Page 10: ...Butane Gasoline Trichloroethylene Vaporized hydrocarbons Hydrogen sulfide Lighter than Air Gas Sensor Location Natural gas methane Near ceiling DO NOT locate directly above appliances where it is sub...

Page 11: ...when the instrument is in the non fault state the contact conditions given on the circuit board next to the terminal are for the non energized state which is identical to the fault state These relay c...

Page 12: ...eases above that of the alarm set point the associated red LED is lit the associated relay changes state and the audio alarm is activated If the dew point increases above that of the alarm set point t...

Page 13: ...is the delay of the GSM 60 staying in alarm condition until after the measured reading has returned past the alarm point by the differential value Example If the alarm point is 10 and the differential...

Page 14: ...S CH 1 CH 3 CH 2 CH 4 O Press Option switch S Press Select switch O No Function L10 A1 vL19 5 vL 40 D 500 O No Function S S L20 A2 23 5 L39 1000 O No Function S Enter Maint Menu Enter Password _ See M...

Page 15: ...ng blocked The system cannot detect a flow fault if only one of the two sampling hoses is blocked 4 7 2 Other Fault Indications Other fault indications are associated with sensor zero and calibration...

Page 16: ...maintenance menu is guarded with a four digit Password The factory default setting of the password is 1270 When a valid numerical password is inserted the user can enter the maintenance menu To enter...

Page 17: ...ess OPTION until the channel to be Calibrated is displayed See Section 5 3 for calibration instructions O S MAINTENANCE MENU Set Alarm1 Alarm 1 Select XX Press OPTION until the gas alarm to be Set is...

Page 18: ...upper portion of display The mV reading Low Cal 0 is displayed in the lower portion of display This is the Low Cal setting usually zero clean air must be supplied to the sensor This reading needs to b...

Page 19: ...to clean uncontaminated air Use a cylinder of 20 9 oxygen to provide a clean air reference if necessary Attach the cylinder to the calibration adapter attach the adapter to the instrument and allow g...

Page 20: ...e are 4 20mA style sensors for dew point or other gases This function is normally performed at the factory and is not usually required to be performed in the field unless a new transmitter is installe...

Page 21: ...cator Press the SELECT switch to lock in the correct indicator ALARM 1 STD is displayed Press the OPTION switch to toggle between STD and DIFF select the correct indicator Press the SELECT switch to l...

Page 22: ...re cursor Locks underscored character and moves cursor to next O Press OPTION until the channel to be Set is displayed Alarm 2 V L 0000 O Press Option S Press Select O S S MAINTENANCE MENU Set Alarm1...

Page 23: ...PTION switch until desired gas is displayed Press the SELECT switch ALARM Delay 0000 is displayed with the underscore cursor under the left digit Press the OPTION switch to change the left digit selec...

Page 24: ...are returned to the Enter Maint Menu display After entering a valid password Press the OPTION switch until Configure Alarms is displayed Press the SELECT switch to enter the Configure Alarms menu Pre...

Page 25: ...tch to set relay indicated On or Off as appropriate Press the SELECT switch to cycle through each of the 5 relays return to Maintenance Menu Relay Failsafes Press OPTION switch until Exit Maint Menu a...

Page 26: ...r digits of the new password have been selected Set New Password is displayed Record the new password without it the maintenance menu cannot be reentered once you exit the Maintenance Menu If the pass...

Page 27: ...strument to properly set operational parameters for each sensor Calibration is the process of setting the instrument up to read accurately when exposed to a target gas This is a two step process A Low...

Page 28: ...the display will move to Hi Cal xx gas calibration xx the level of gas to be used for calibration The mV reading is shown in the upper right hand corner of the display Apply calibration gas to sensor...

Page 29: ...an uncontaminated air Use a cylinder of 20 9 oxygen to provide a clean air reference if necessary Attach the cylinder to the calibration adapter attach the adapter to the instrument and allow gas to f...

Page 30: ...tained medical air systems this orifice should not clog However if difficulty is experienced in maintaining flow rate examine this orifice replace it if necessary Orifice location will depend on instr...

Page 31: ...lug 3 position 65057 012 Terminal plug 4 position 65057 010 Terminal plug 2 position 73089 004 Filter Assembly Particulate 04018 128 Pump Replacement for instrument s n 118 and above 04018 118 Pump Re...

Page 32: ...he presence of CO and to facilitate the assessment of the degree of danger that he or she is exposed to Based upon knowledge of the effects of CO the Occupational Safety and Health Authority OSHA has...

Page 33: ...omopropene 9 30 Chlorofluoromethane Freon 22 12 45 1 Butanethiol 9 14 Chloroform 11 37 1 Butene 9 58 Chlorotrifluoromethane Freon 13 12 91 1 Butyne 10 18 Chrysene 7 59 2 3 Butadione 9 23 Cresol 8 14 2...

Page 34: ...phenyl 7 95 Hydrogen chloride 12 74 Dipropyl amine 7 84 Hydrogen cyanide 13 91 Dipropyl sulfide 8 30 Hydrogen fluoride 15 77 Durene 8 03 Hydrogen iodide 10 38 m Dichlorobenzene 9 12 Hydrogen selenide...

Page 35: ...ormate 10 54 Methyl isobutyrate 9 98 Propylene 9 73 Methyl isocyanate 10 67 Propylene dichloride 10 87 Methyl isopropyl ketone 9 32 Propylene imine 9 00 Methyl isothiocyanate 9 25 Propylene oxide 10 2...

Page 36: ...l Name IP eV Chemical Name IP eV T continued W Trimethyl amine 7 82 Water 12 59 Tripropyl amine 7 23 X V 2 4 Xylidine 7 65 o Vinyl toluene 8 20 m Xylene 8 56 Valeraldehyde 9 82 o Xylene 8 56 Valeric a...

Page 37: ...he inspection by the Seller confirms that the product is defective it will be repaired or replaced at no charge within the stated limitations and returned prepaid to any location in the United States...

Page 38: ...the Service Request Form assists in the expedient service and return of your unit and failure to provide this information can result in processing delays ENMET charges a one hour minimum billing for a...

Page 39: ...it Appears on Card RETURN SHIPPING METHOD UPS Ground UPS 3 Day Select UPS Next Day Air UPS ND Air Saver UPS 2 Day Air UPS Account FedEx Ground FedEx Air Express Saver FedEx Air Overnight Std FedEx Air...

Reviews: