TROUBLESHOOTING 

ERROR 
MESSAGE

 

PROBLEM 

SOLUTION 

 

L

o

 

 
Temperature taken is 
lower than 90.0°F(32.0°C)

 

Turn off, wait one minute and take a 
new temperature via close contact and 
sufficient rest. 

 

H

i

 

 
Temperature taken is  
higher than 109.9°F(42.9°C) 

Turn off, wait one minute and take a 
new temperature via close contact and 
sufficient rest. 

 
 

Err 

 
The system is not functioning  
properly. 

Unload the battery, wait for 1 minute 
and repower it.  If the message 
reappears, contact the retailer for 
service. 

 

 

Dead battery: Battery icon is 
flashing, can’t measure. 

 
Replace the battery. 
 

 

How To Calibrate Your Thermometer: 
 

The thermometer is initially calibrated at the time of manufacturing. If the thermometer is 
used  according  to  the  manual,  periodic  readjustment  is  not  required.  However,  we 
recommend conducting a quick calibration to confirm that your unit didn't get damaged during 
shipment.  A  quick  way  to  calibrate  your  unit  is  by  doing  the  water  test.  Dip  the  tip  of  the 
thermometer  into  a  cup  of  hot  water.  Wait  for  your  thermometer  to  show  'Hi'  or  a  high 
temperature reading number on its LCD display (for example it shows 38 °C or 100 °F). Then 
let it sit for a minute or two then dip it into a cup of cold water. Your display should show 'Lo' 
(meaning low temperature not low battery) or a low temperature reading number this time 
(for example it shows 33 °C or 91°F). Leave it at room temperature for 30 minutes and try using 
it again, and if it does not show ‘Hi’ or ‘Lo’  it will mean you have successfully checked your 
thermometer and good to go! If you find that your reading is very low and/or not changing on 
the LCD display at all, please get in touch with us through email with your Amazon Order ID.  
 

 SYMBOL EXPLANATION 

 

 

 

Direct Current 

 

 

 

Batch Code 

 

 

 

 

Type BF Applied Part 

       

 

 

Manufacturer  

  

 

Consult 
Accompanying  
Documents 

 

 

Storage and Transportation 
Temperature Limit: -4 

o

F ~ 131

o

F (-

20

 o

C ~ 55

o

C) 

 

CLEANING & DISINFECTION 

Wipe the thermometer with a soft clean cloth. For stubborn stains, wipe the  thermometer 
with a cloth that has been dampened in water or in neutral detergent solution and then wring 
thoroughly. Finish by wiping with a soft dry cloth. For disinfection, 75% Ethanol or Isopropyl 
alcohol can be used. Observe the following to prevent damage to the thermometer: 

 

Do  not  use  benzene,  thinner,  gasoline  or  other  strong  solvents  to  clean  the 
thermometer. 

 

Do  not  attempt  to  disinfect  the  sensing  section  (tip)  of  the  thermometer  by 
immersing in alcohol or in hot water (water over 122oF/50oC). 

 

Do not use ultrasonic washing to clean the thermometer. 

 

 

o

F/

o

C SWITCHABLE 

Temperature readings are available in the Fahrenheit or Celsius scale (

o

F/

o

C; located in the 

upper right corner of LCD). With the unit off, press and hold On/Off Button for approximately 
2 seconds to change the current setting.

 

 
 
 
 

SPECIFICATIONS: 

Type: 

Digital Thermometer (Not Predictive) 

Measure range: 

90.0

o

F-109.9

o

F (32.0

o

C-42.9 

o

C) (

o

/

o

F chosen by manufacturer) 

Accuracy: 

   0.2

o

F (   0.1

o

C) during 95.9

o

F ~ 

107.6

o

F (35.5

o

C ~ 42.0

o

C) at 64.4

o

F ~ 

82.4

o

F (18

o

C ~ 28

o

C) ambient 

operating range    0.4

o

F (   0.2

o

C) for 

other measuring and ambient 
operating range 

Operating mode: 

Direct Mode 

Display: 

Liquid crystal display, 3 ½  digits 

Memory: 

For storing the last measured value  

Battery: 

One 1.5 V DC. Button battery (size 
LR41 or SR41, UCC 392) 

Battery life: 

Approx. 200 hours of continuous 
operation of 1 year with 3 
measurements per day 

Dimension: 

13.9 cm x 2.2 cm x 1.2 cm (L x W x H)  

Weight: 

Approx. 12 grams including battery 

Expected service 
life: 

Three years 

 
Ambient operating 
range: 

Temperature: 41

o

F ~ 104

o

F (5

o

C ~ 

40

o

C) 

Relative humidity: 15% ~ 95% RH 
Atmospheric Pressure: 700hPa ~ 
1060hPa 

Storage and 
transportation 
condition: 

Temperature: -4

o

F ~ 131

o

F (-20

o

C ~ 

55

o

C) 

Relative humidity: 15% ~ 95 % RH 
Atmospheric Pressure: 700hPa ~ 
1060hPa 

Ingress Protection 
Rating: 

IP 27 

Classification: 

Type BF    

 
BATTERY REPLACEMENT 

1.

 

Replace battery when   appears in the lower 
right corner of LCD display. 

2.

 

Pull battery cover off as shown in Figure 3. 

3.

 

Gently pull out plastic circuit board with 
battery chamber approximately 1 cm (slightly 
less than ½”.) (See Figure 4) 

4.

 

Use pointed object such as a pen to remove 
old battery. Discard battery lawfully. Replace 
with new 1.5 V DC button type LR41 or SR41, 
UCC392, or equivalent. Be sure battery is 
installed with  
“+” polarity facing up. (See Figure 5) 

5.

 

Slide battery chamber back into place and 
attach cover. 
 

      

      

    

  

 

 
 
 