TROUBLESHOOTING
ERROR
MESSAGE
PROBLEM
SOLUTION
L
o
Temperature taken is
lower than 90.0°F(32.0°C)
Turn off, wait one minute and take a
new temperature via close contact and
sufficient rest.
H
i
Temperature taken is
higher than 109.9°F(42.9°C)
Turn off, wait one minute and take a
new temperature via close contact and
sufficient rest.
Err
The system is not functioning
properly.
Unload the battery, wait for 1 minute
and repower it. If the message
reappears, contact the retailer for
service.
Dead battery: Battery icon is
flashing, can’t measure.
Replace the battery.
How To Calibrate Your Thermometer:
The thermometer is initially calibrated at the time of manufacturing. If the thermometer is
used according to the manual, periodic readjustment is not required. However, we
recommend conducting a quick calibration to confirm that your unit didn't get damaged during
shipment. A quick way to calibrate your unit is by doing the water test. Dip the tip of the
thermometer into a cup of hot water. Wait for your thermometer to show 'Hi' or a high
temperature reading number on its LCD display (for example it shows 38 °C or 100 °F). Then
let it sit for a minute or two then dip it into a cup of cold water. Your display should show 'Lo'
(meaning low temperature not low battery) or a low temperature reading number this time
(for example it shows 33 °C or 91°F). Leave it at room temperature for 30 minutes and try using
it again, and if it does not show ‘Hi’ or ‘Lo’ it will mean you have successfully checked your
thermometer and good to go! If you find that your reading is very low and/or not changing on
the LCD display at all, please get in touch with us through email with your Amazon Order ID.
SYMBOL EXPLANATION
Direct Current
Batch Code
Type BF Applied Part
Manufacturer
Consult
Accompanying
Documents
Storage and Transportation
Temperature Limit: -4
o
F ~ 131
o
F (-
20
o
C ~ 55
o
C)
CLEANING & DISINFECTION
Wipe the thermometer with a soft clean cloth. For stubborn stains, wipe the thermometer
with a cloth that has been dampened in water or in neutral detergent solution and then wring
thoroughly. Finish by wiping with a soft dry cloth. For disinfection, 75% Ethanol or Isopropyl
alcohol can be used. Observe the following to prevent damage to the thermometer:
Do not use benzene, thinner, gasoline or other strong solvents to clean the
thermometer.
Do not attempt to disinfect the sensing section (tip) of the thermometer by
immersing in alcohol or in hot water (water over 122oF/50oC).
Do not use ultrasonic washing to clean the thermometer.
o
F/
o
C SWITCHABLE
Temperature readings are available in the Fahrenheit or Celsius scale (
o
F/
o
C; located in the
upper right corner of LCD). With the unit off, press and hold On/Off Button for approximately
2 seconds to change the current setting.
SPECIFICATIONS:
Type:
Digital Thermometer (Not Predictive)
Measure range:
90.0
o
F-109.9
o
F (32.0
o
C-42.9
o
C) (
o
C
/
o
F chosen by manufacturer)
Accuracy:
0.2
o
F ( 0.1
o
C) during 95.9
o
F ~
107.6
o
F (35.5
o
C ~ 42.0
o
C) at 64.4
o
F ~
82.4
o
F (18
o
C ~ 28
o
C) ambient
operating range 0.4
o
F ( 0.2
o
C) for
other measuring and ambient
operating range
Operating mode:
Direct Mode
Display:
Liquid crystal display, 3 ½ digits
Memory:
For storing the last measured value
Battery:
One 1.5 V DC. Button battery (size
LR41 or SR41, UCC 392)
Battery life:
Approx. 200 hours of continuous
operation of 1 year with 3
measurements per day
Dimension:
13.9 cm x 2.2 cm x 1.2 cm (L x W x H)
Weight:
Approx. 12 grams including battery
Expected service
life:
Three years
Ambient operating
range:
Temperature: 41
o
F ~ 104
o
F (5
o
C ~
40
o
C)
Relative humidity: 15% ~ 95% RH
Atmospheric Pressure: 700hPa ~
1060hPa
Storage and
transportation
condition:
Temperature: -4
o
F ~ 131
o
F (-20
o
C ~
55
o
C)
Relative humidity: 15% ~ 95 % RH
Atmospheric Pressure: 700hPa ~
1060hPa
Ingress Protection
Rating:
IP 27
Classification:
Type BF
BATTERY REPLACEMENT
1.
Replace battery when appears in the lower
right corner of LCD display.
2.
Pull battery cover off as shown in Figure 3.
3.
Gently pull out plastic circuit board with
battery chamber approximately 1 cm (slightly
less than ½”.) (See Figure 4)
4.
Use pointed object such as a pen to remove
old battery. Discard battery lawfully. Replace
with new 1.5 V DC button type LR41 or SR41,
UCC392, or equivalent. Be sure battery is
installed with
“+” polarity facing up. (See Figure 5)
5.
Slide battery chamber back into place and
attach cover.
All rights reserved.