HEMACOOL
TM
Model HMC-MIL-1 Operations Manual August 2003
Energy Storage Technologies, Inc.
Dayton, Ohio 45459
10
C. STARTING THE HemaCool™ :
CAUTION: Do not turn on the HemaCool™
until you have read the entire user’s
manual. Failure to understand the meaning of the LED readouts could result in loss
of stored blood, blood products, or medical supplies and possibly do damage to the
unit.
Fig. 9. HemaCool™
Display
1.
Select the temperature
you want to operate the unit with the toggle switch on the
power panel (
Fig. 8
number
1
).
The COOL (+4.0 ) control point is ONLY used for storing materials that must be
stored above freezing. The FREEZE (-22° C) control point is ONLY used for
materials to be stored frozen. (If the ambient temperature is more than +30°C
[+86°F], the device may not have the capacity to hold the payload at FREEZE
temperature below -20°C [-4°F]. See specifications at the end of this document.)