19
CHAPTER 3
POWER REQUIREMENTS
Power is supplied by a battery pack consisting of four D-size Ni-MH rechargeable
batteries. A 120 –240 Volt AC charger having a 12 Volt DC/ 1.5 A. output is
supplied with the instrument. The battery charger will charge the 9,000 mAH
batteries in about six hours.
Battery life is approximately 4-6 hours of continuous operation. When the battery
voltage drops to 4.0 volts, a “battery low” warning will appear on the display. A
few minutes later the instrument will automatically turn off to preserve the
remaining battery power for the sensor biases.
•
The CO, NO, and SO2 sensors needs a tiny amount of electrical power
even when the analyzer is off.
•
Do not allow the batteries to discharge completely. When not in daily use,
remember to charge your analyzer periodically.
•
If the analyzer has been without power for a long time, you may need to wait
a few hours after recharging the battery for the sensors to be restored.
You can check the condition of the battery at any time:
1.
Press the
DISPLAY DATA
key and observe the battery icon, located at the bottom
of the display.
Battery Full
Battery Empty
Battery Charger Connected
Fast-Charge Mode
2.
Press the
LEFT ARROW
and
RIGHT ARROW
keys together and observe the
current battery voltage, listed as
BAT
. When the unit is operating on its
batteries, the voltage displayed will vary from an initial 5.2–5.4 volts (fully
charged), dropping slowly to 4.2 volts (batteries nearly empty).
Summary of Contents for 700
Page 2: ...2 ENERAC 700 MANUAL EDITION 4 M700 FIRMWARE 3 0 M 700 HARDWARE H ...
Page 4: ...4 APPENDIXC EPA TEST METHODS 70 ...
Page 14: ...14 ...
Page 18: ...18 ...
Page 20: ...20 ...
Page 49: ...49 ...
Page 63: ...63 ...
Page 67: ...67 ...