Troubleshooting
Turbimax CUE25 / CUE26
18
Hauser
7
Troubleshooting
7.1
Troubleshooting instructions
The instrument routinely performs self-diagnostics and provides you with two levels of specific
diagnostic information.
• Battery low indicator
• System error messages
Refer to the sections "Battery low indicator" and "System error messages" to localize and rectify a
warning condition or an error.
7.2
System error messages
Normally, the cause of an error message is external to the instrument. The following table lists the
error messages and their associated meanings:
7.3
Battery low indicator
7.4
Spare parts
Error message
Typical cause
Tests and / or corrective measures
"Err" icon on
Internal error
Contact your responsible service.
"Err" icon on; display reads
"CAL"
Bad standards or standards in
wrong order.
Check the standards and recalibrate the instrument. See
section "Instrument calibration" for detailed
information.
Warning
Possible cause
Tests and / or corrective measures
Flashing battery icon
Batteries are low
Batteries need to be replaced but the readings displayed
are still accurate.
Solid battery icon
Power level too low
Readings may be out of specification. Replace the
batteries as soon as possible to ensure that the Turbimax
will continue to function properly.
If the batteries get too low to perform, the instrument
will turn off and may not turn on until the batteries have
been replaced. See section "Replacing the batteries" for
more information.
Designation and content
Order number spare part kit
Rechargeable battery kit
Please contact the responsible
service.