background image

Operation

Turbimax CUE25 / CUE26

14

Hauser

Step three:

1.

The instrument prompts for the 0.02 NTU standard with a flashing "0.02".

2.

Insert the 0.02 NTU standard into the optical well and press down until the cuvette snaps fully 
into the instrument. Align the indexing ring with the arrow on the instrument.

3.

Press and hold the 

F

 key and fine-tune the standard to the lowest value on the display by 

slowly rotating it back and forth about 5°.

4.

Release the 

F

 key to initiate calibration. The instrument will show a 30 second count down.

5.

When the instrument has completed calibration on the 0.02 NTU standard, it returns to the 
read mode and is ready to read.

!

Note! 
• If you wish to exit the calibration mode, you may do so at the end of any step simply by pressing 

the 

Ï

 key. The instrument will store any of the values calibrated prior to exiting.

• You can cycle through the required calibration points (0.02 NTU, 10 NTU, 1000 NTU) by 

pressing either the 

V

 or 

W

 keys to individually calibrate with a particular calibration standard.

• The required calibration values are set in software and cannot be changed.

Calibration error

If the calibration failed, the instrument will display the error shown below. 

Calibration failure can have the following reasons:
• Bad calibration standards
• Wrong order of calibration standards

Either check the standards and recalibrate or restore the factory calibration by performing the 
following operation:

1.

Push and hold the 

V

 key.

2.

Push and release the 

F

 key and then release the 

V

 key.

3.

 The instrument will turn itself off. When it is turned on again, the error screen will be gone.

!

Note! 
After restoring factory calibration, the instrument may operate with reduced accuracy until a new 
calibration can be performed.

a0003815

NTU

FNU

-rd-

a0003817

NTU

FNU

CAL

ERR

Summary of Contents for Turbimax CUE25

Page 1: ...Operating Instructions Turbimax CUE25 CUE26 Handheld turbidimeter for field measurement AUTHORIZED DISTRIBUTOR InstrumentsAndControl com Houston Texas USA sales InstrumentsAndControl com 281 609 7170...

Page 2: ...ere Here you can find the steps for installation of the batteries Operation Page 10 Page 11 ff Use this section to familiarize yourself with the operation of the device This section provides informati...

Page 3: ...Installation instructions 9 3 5 Post installation check 9 4 Operation 10 4 1 Display and operating elements 10 4 2 Key assignment 10 4 3 Routine operation 11 4 4 Instrument calibration 12 5 Maintenan...

Page 4: ...int check all the connections for correctness Ensure that electrical cables and hose connections are not damaged Do not operate damaged products and secure them against unintentional commissioning Mar...

Page 5: ...cause damage to the instrument if ignored Note This symbol indicates important items of information Direct Current DC A terminal at which DC is applied or through which DC flows Alternating Current A...

Page 6: ...0 02 NTU standard 10 0 NTU standard 1000 NTU standard 2 Empty cuvettes and Kimwipes 1 Operating Instructions BA397C 07 en 1 Instruction card If you have any questions please contact your supplier or...

Page 7: ...pplier about damage to the delivery contents Keep the damaged products until the matter has been settled Check that the scope of delivery is complete and agrees with your order and the shipping docume...

Page 8: ...llation Turbimax CUE25 CUE26 8 Endress Hauser 3 3 Installation conditions a0004367 Fig 3 Dimensions CUE25 CUE26 a0004368 Fig 4 Dimensions carry case 70 2 75 165 6 50 48 1 88 mm inch 250 10 216 8 50 mm...

Page 9: ...ce 2 Remove the two screws pos 1 securing the battery cover pos 2 Lift the cover off 3 Place the supplied four AAA alkaline batteries pos 3 into the battery compartment pos 4 ensuring the correct pola...

Page 10: ...ay of turbidity levels and user guidance in calibration routine 4 2 Key assignment a0003588 Fig 7 Operating elements 1 key used to turn the Turbimax on or off 2 V W keys used to set numerical values a...

Page 11: ...he cuvette with the black light shield cap Ensure that the outside of the cuvette is dry clean and free from smudges1 7 Place the cuvette into the instrument and press it down until it snaps fully int...

Page 12: ...d immediately for calibration and then be disposed of If the 10 0 NTU standard is stored in the glass cuvette for extended periods of time it should be used as a reference standard only 4 4 2 Indexing...

Page 13: ...5 4 Release the F key to initiate the calibration The display will show a 30 second count down 5 When the instrument has completed calibration on the 1000 NTU it will briefly display 1000 to indicate...

Page 14: ...l store any of the values calibrated prior to exiting You can cycle through the required calibration points 0 02 NTU 10 NTU 1000 NTU by pressing either the V or W keys to individually calibrate with a...

Page 15: ...or scratches To clean a cuvette proceed as follows 1 Clean the interior and exterior with a detergent solution 2 Rinse the cuvette thoroughly 8 to 10 times with distilled or deionized water to elimina...

Page 16: ...e it upside down on a soft surface 2 Remove the two screws pos 1 securing the battery cover pos 2 Lift the cover off 3 Remove the old batteries 4 Place four new AAA batteries pos 3 into the battery co...

Page 17: ...ssories Endress Hauser 17 6 Accessories 6 1 Calibration solutions Calibration kit CUE25 CUE26 full range 0 02 NTU 10 0 NTU 1000 NTU Order no 51518582 6 2 Cuvettes Sample cuvettes CUE25 CUE26 incl caps...

Page 18: ...ternal error Contact your responsible service Err icon on display reads CAL Bad standards or standards in wrong order Check the standards and recalibrate the instrument See section Instrument calibrat...

Page 19: ...red please return it cleaned to the sales center responsible Please use the original packaging if possible 7 6 Disposal The device contains electronic components and must therefore be disposed of in a...

Page 20: ...ime 14 s Reference temperature 25 C 77 F Resolution 0 01 NTU below 100 NTU 0 1 NTU in the range 100 0 to 999 9 NTU 1 0 NTU in the range 1000 to 1100 NTU Maximum measured error 0 to 500 NTU 500 to 1100...

Page 21: ...1 875 x 2 75 x 6 50 Instrument in case H x W x L 63 x 250 x 216 mm 2 50 x 10 x 8 50 Weight 1 2 kg 2 7 lbs Materials Instrument housing ABS injection molded Sample cuvette Borosilicate glass Carry cas...

Page 22: ...Incoming acceptance 7 Input 20 Installation 4 7 9 Instrument calibration 12 K Key assignment 10 M Maintenance 15 Batteries 16 Cuvette 15 Measuring system 7 Mechanical construction 21 N Nameplate 6 O...

Page 23: ...Turbimax CUE25 CUE26 Endress Hauser 23...

Page 24: ...www endress com worldwide BA397C 07 en 07 06 FM SGML 6 0 DT 71001162...

Reviews: