Tophit H CPS471/CPS471D
16
Hauser
Certificates and approvals
Ex approval FM/CSA
• FM
Cl. I, Div. 1, Groups A, B, C, D, associated apparatus Mycom S 153-O/-P or Liquiline M CM42-NP/-PP
• CSA
Cl. I, Div. 1, Groups A, B, C, D, associated apparatus Mycom S 153-S or Liquiline M CM42-NS/-PS
Ex approval ATEX
Device group II, Category 1G
Explosion protection EEx ia IIC T3/T4/T6
Sanitary compatibility EHEDG
Sterilisability validated acc. to EHEDG test criteria (TNO Report V3640 RE)
Original of report summary
"At the request of Hauser Conducta, Postfach 100154, D-70826 Gerlingen, Germany the in-place
cleanability of the ISFET-Sensor TopHit CPS 471 was assessed according to the test procedure of the
European Hygienic Engineering & Design Group (EHEDG).
The test results show that the ISFET-Sensor TopHit CPS 471, including the seals, is cleanable in-place at least
as well as the reference pipe. The tests were conducted five times on one test object. The results of the tests
are comparable with each other. The ISFET-Sensor TopHit CPS 471 complies with the hygienic criteria of the
Machinery Directive 98/37/EC, annex 1 (additional essential health and safety requirements for certain
categories of machinery) section 2.1 (agri-foodstuffs machinery), the hygienic requirements of EN 1672 - part
2 and with the hygienic equipment design criteria of the EHEDG.
The test results obtained are representative of the ISFET-Sensor TopHit type CPS 471 and type CPS 441."
• Materials in contact with medium are FDA certified
• Certified acc. to the 3-A standard no. 74-02
Accessories
Transmitters
• Liquiline M CM42
Modular two-wire transmitter, stainless steel or plastic, field or panel instrument,
various Ex approvals (ATEX, FM, CSA, Nepsi, TIIS),
HART, PROFIBUS or FOUNDATION Fieldbus available
Ordering acc. to product structure, see Technical Information (TI381C/07/en)
• Liquisys M CPM223/253
Transmitter for pH and redox, field or panel-mounted housing,
HART or PROFIBUS available
Ordering acc. to product structure, see Technical Information (TI194C/07/en)
• Mycom S CPM153
Transmitter for pH and redox, one or two channel version, Ex or Non-Ex,
HART or PROFIBUS available
Ordering acc. to product structure, see Technical Information (TI233C/07/en)
Fully automatic measuring
systems
Topcal S CPC310
• Fully automatic measuring, cleaning and calibration system; Ex or Non-Ex
• in-situ cleaning and calibration, automatic sensor monitoring
• Ordering acc. to product structure, Technical Information TI404C/07/de
Topclean S CPC30
• Fully automatic measuring and cleaning system; Ex or Non-Ex
• in-situ cleaning, automatic sensor monitoring
• Ordering acc. to product structure, see Technical Information TI235C/07/en
Service tool
Memocheck Plus CYP01D, Memocheck CYP02D
• Tool for the qualification of measuring chains
• Service tool for quick, on-site checks of measuring systems with Memosens technology
• Verification of data transmission
• Ordering acc. to product structure, KA399C/07/a2