background image

Tophit H CPS471/CPS471D

Hauser

5

Memosens (CPS471D)

Maximum process safety

The inductive and non-contacting measured value transmission of Memosens guarantees maximum process 
safety and offers the following benefits:

• All problems caused by moisture are eliminated.

– The plug-in connection is free from corrosion.
– Measured value distortion from moisture is not possible.
– The plug-in system can even be connected under water.

• The transmitter is galvanically decoupled from the medium. The result: No more need to ask about 

"symmetrically high-impedance" or "unsymmetrical" (for pH/ORP measurement) or an impedance 
converter.

• EMC safety is guaranteed by screening measures for the digital measured value transmission.

Data safety through digital data transfer

The Memosens technology digitalizes the measured value in the sensor and transfers it to the transmitter via a 
contactless connection. The result:

• An automatic error message is generated if the sensor fails or the connection between sensor and transmitter 

is interrupted.

• The availability of the measuring point is dramatically increased by immediate error detection.
• The digital signals are suitable for application in hazardous areas; the integrated electronics are intrinsically 

safe.

Easy handling

Sensors with Memosens technology have integrated electronics that allow for saving calibration data and 
further information such as total hours of operation and operating hours under extreme measuring conditions. 
When the sensor is mounted, the calibration data are automatically transferred to the transmitter and used to 
calculate the current measured value. Storing the calibration data in the sensor allows for calibration and 
adjustment away from the measuring point. The result:

• Sensors can be calibrated unter optimum external conditions in the measuring lab. Wind and weather do 

neither affect the calibration quality nor the operator.

• The measuring point availability is dramatically increased by the quick and easy replacement of precalibrated 

sensors.

• The transmitter does not need to be installed close to the measuring point but can be placed in the control 

room.

• Maintenance intervals can be defined based on all stored sensor load and calibration data and predictive 

maintenance is possible.

• The sensor history can be documented on external data carriers and evaluation programs at any time. Thus, 

the current application of the sensors can be made to depend on their previous history.

Communication with the transmitter

Always connect digital sensors with Memosens technology to a transmitter with Memosens technology. Data 
transmission to a transmitter for analog sensors is not possible.

The sensor is connected to the cable connection (CYK10) without contact. The power and data are transferred 
inductively
Once connected to the transmitter, the data saved in the sensor are read digitally. You can call up these data 
using the corresponding DIAG menu.

Data that digital sensors save include the following: 
• Manufacturer data

– Serial number
– Order code
– Date of manufacture

• Calibration data

– Calibration date
– Calibration values
– Number of calibrations
– Serial number of the transmitter used to perform the last calibration

• Operational data

– Date of commissioning
– Hours of operation under extreme conditions
– Number of sterilizations
– Data for sensor monitoring.

Summary of Contents for Tophit CPS471

Page 1: ...ference system poisoning resistant polyacrylamide free gel Application possible at low temperatures Short response time Constantly high accuracy Sterilisable and autoclavable Longer calibration interv...

Page 2: ...table sensor circuits use the dependence on the ion selective gate potential to create an output signal proportional to the concentration of the ion type pH selective IsFET The gate isolator serves as...

Page 3: ...lues 0 09 and 13 86 a0003867 en Fig 3 Comparison of acid and alkaline errors Measurement stability and sensor response time The ISFET response times are very short over the whole temperature range Wit...

Page 4: ...H value can be worked out from this equation isothermic curve potential change per pH value at a defined temperature The isothermic curves of the ISFET sensor are very close to the theoretical values...

Page 5: ...mitter and used to calculate the current measured value Storing the calibration data in the sensor allows for calibration and adjustment away from the measuring point The result Sensors can be calibra...

Page 6: ...sembly e g Cleanfit P CPA471 CPA450 with digital sensor only There are additional accessories available depending on the application Topclean S CPC30 or Topcal S CPC310 automatic cleaning system Exten...

Page 7: ...tomatic measurement cleaning and calibration system CIP cleaning The sensor built into the retractable assembly is automatically moved out of the medium before cleaning In the rinse chamber of the ret...

Page 8: ...mmersion assembly Dipfit W CPA111 3 Special measuring cable CPK12 or CYK10 4 Transmitter Liquiline Pharmaceuticals and biotechnology a0003877 Fig 8 Measuring system for pharmaceutical and biotechnolog...

Page 9: ...ote The cable cors Yellow and White are connected on the sensor side Make sure you comply with the instructions for connecting the sensor wiring diagram in the Operating Instructions of the transmitte...

Page 10: ...s Longer dry storage of the sensor up to 30 minutes Installation Installation angle ISFET sensors can be installed in any position as there is no liquid internal lead However in case of an overhead in...

Page 11: ...he rotable plug in head When installing the sensor in an assembly use the engraved serial number on the connection head for correct sensor orientation The serial number is always located in the same p...

Page 12: ...alibration and operation Normal environment light does not influence the measurement Process Medium temperature depending on pH At high temperatures over a long period of time alkalis irreversibly des...

Page 13: ...en Fig 14 Pressure and temperature Caution Danger of damage to the sensor Never use the Tophit for applications outside the given specifications Recommended cleaning Depending on the degree of polluti...

Page 14: ...acc to DIN IEC 751 Plug in head CPS471 ESB TOP68 rotatable CPS471D Memosens rotatable Diaphragm Ceramics sterilizable a0003903 Fig 15 Tophit CPS471 depending on the sensor version a0003904 Fig 16 Sens...

Page 15: ...l EPDM Application range pH 0 to 14 15 to 135 C 5 to 275 F For Ex and Non Ex applications Product structure CPS471 Product structure CPS471D Shaft length 2 120 mm 4 72 in 4 225 mm 8 86 in 5 360 mm 14...

Page 16: ...obtained are representative of the ISFET Sensor TopHit type CPS 471 and type CPS 441 Materials in contact with medium are FDA certified Certified acc to the 3 A standard no 74 02 Accessories Transmit...

Page 17: ...iable process termination parts in contact with medium of PP PEEK or PVDF Technical Information TI345C 07 en Cleanfit H CPA475 Retractable assembly for tank and pipe installation under sterile conditi...

Page 18: ...sheath max 130 C 266 F HC Cable length 15 m 49 23 ft TPE sheath max 130 C 266 F HD Cable length 20 m 65 64 ft TPE sheath max 130 C 266 F HF Cable length 5 to 20 m 16 41 to 65 64 ft TPE sheath max 130...

Page 19: ...of e g Memosens CUS31 CUS41 2 wires twisted pair with shield and PVC sheath 2 x 2 x 0 5 mm2 shield Sold by the meter order no 51502543 Junction boxes Junction box VBA With 10 high impedance terminals...

Page 20: ...TI283C 07 en 04 06 51506685 Printed in Germany FM SGML 6 0 DT...

Reviews: