Appendix A: Specifications
Table A–1
SS500XP H
2
O analyzer specifications
Performance
Typical Measurement Ranges
1
1. Consult factory for alternative ranges.
40 to 422 ppmv (2 to 20 lbs/MMscf)
40 to 1055 ppmv (2 to 50 lbs/MMscf)
40 to 2110 ppmv (2 to 100 lbs/MMscf)
Application Data
Environmental Temperature Range
-20 to 50 °C (-4 to 122 °F)
Environmental Relative Humidity
80% for temperatures up to 31°C MAX
Altitude
Up to 2000 m (6,550 ft.)
Maximum Cell Pressure
10 PSIG (70kPaG)
Sample Cell Pressure Range
700 to 1400 mbara
Sample Flow Rate
0.5 to 1.0 LPM (1 to 2 scfh) typical
Recommended Validation
Bureau of Mines chilled mirror, portable TDL or binary
gas with methane background
Electrical & Communications
Input Voltages
2
2. Supply voltage not to exceed
10% of nominal. Transient over-voltages according to
Overvoltage Category II.
100 to 240 VAC, 50/60 Hz (single phase only)
18 to 24 VDC -
Optional
Contact Rating
(Inductive Load)
250V, 3A N.O. contact, 1.5A N.C. contact
24V, 1A N.O. and N.C. contact
Communication
Analog
: (2) 4–20 mA Isolated, 1200 ohms @ 24 VDC
max load
Serial
: RS-232C
Protocol
: Modbus Gould RTU, Daniel RTU or ASCII
Digital Output
(2); General Fault and Concentration/Assignable Alarm
LCD Display
Concentration, cell pressure, temperature alarms &
diagnostics
Physical Specifications
Size
565 mm H x 413 mm W x 252 mm D
(17.27 in. H x 14.59 in. W x 9.94 in. D)
Weight
Approx. 46.8 Kg (103 lbs)
Sample Cell Construction
316L Series Polished Stainless Steel
Area Classification
Certifications
CSA Class I, Division 1 Groups B, C and D;
T4; -20°C ≤ Tamb ≤ +50°C, Enclosure Type 4