Troubleshooting & Maintenance
Hardware Manual
B
–
5
Tools and Supplies:
•
Lens cleaning cloth (Cole Parmer
®
EW-33677-00 TEXWIPE
®
Alphawipe
®
Low-Particulate Clean Room Wipes or equivalent)
•
Reagent-grade isopropyl alcohol (ColeParmer
®
EW-88361-80 or
equivalent)
•
Small drop dispenser bottle (Nalgene
®
2414 FEP Drop Dispenser
Bottle or equivalent)
•
Acetone-impenetrable gloves (North NOR CE412W Nitrile Chemsoft™
CE Cleanroom Gloves or equivalent)
•
Hemostat (Fisherbrand™ 13-812-24 Rochester-Pean Serrated
Forceps)
•
Bulb blower or dry compressed air/nitrogen
•
Torque wrench
•
Permanent ink marker
•
Non-outgassing grease
•
Flashlight
To clean the mirrors:
1.
Power down the analyzer following the procedure outlined in
“To
power down the analyzer”
in the Firmware Manual.
2.
Isolate the sample conditioning system
(
SCS) from the process
sample tap.
The sample cell assembly contains a low-power, 20 mW MAX, CW
Class 3b invisible laser with a wavelength between 800-3000 nm.
Never open the sample cell flanges or the optical assembly unless
the power is turned off.
The sample cell assembly contains a low-power, 20 mW MAX, CW
Class 3b invisible laser with a wavelength between 800-3000 nm.
Never open the sample cell flanges or the optical assembly unless
the power is turned off.
All valves, regulators, switches, etc. should be operated in
accordance with site lock-out/tag-out procedures.
Summary of Contents for SpectraSensors SS2100r
Page 1: ...SS2100r H2S Analyzer Rack Mount Hardware Manual P N 4900002189 rev D...
Page 2: ......
Page 4: ......
Page 8: ...SS2100r H2S Analyzer iv 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 10: ...SS2100r H2S Analyzer vi 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 34: ...SS2100r H2S Analyzer 2 8 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 44: ...SS2100r H2S Analyzer A 4 4900002189 rev D 3 7 18 THIS PAGE INTENTIONALLY LEFT BLANK...