Electrical connection
Prosonic S FMU90 PROFIBUS DP
26
Hauser
Terminals for temperature input
Terminal area D
Terminals 83, 84, 85:
• Pt100
• Omnigrad S TR61 ( Hauser)
Other elements on the terminal areas
Terminal area A
•
Display
Connection of the display or the remote display and operating module
•
Service
Service interface; for connection of a PC/notebook via Commubox FXA291
•
Write protection switch: locks the device to prevent modifications to the configuration.
6.3
Special connection instructions
6.3.1
Connecting the power supply
L
CAUTION
To guarantee electrical safety:
‣
Fix the power supply cables securely in place when installing so they are permanently
connected to the electrical installations for the building.
‣
When connecting to the public mains, install a mains switch for the device such that it
is within easy reach of the device. Mark the power switch as a disconnector for the
device (IEC/EN61010).
‣
For the 90-253VAC version: connect potential equalization
‣
Switch off the supply voltage before connecting.
Connection of the power supply in the polycarbonate field housing
Valid for:
Order code 030 (housing, material)
Option 1 (PC field mounting, IP66 NEMA4x)