Commissioning
Prosonic M FMU44 HART
26
Hauser
Parameters that need to be configured when the device is switched on for the first time:
•
Language
Select the language of the display.
•
Length unit
Select the unit of length in which the distance is measured.
•
Basic setup
A measured value is visible that does not yet display the level in the vessel
Perform the basic setup
9.3
Connecting via FieldCare
Commissioning via FieldCare is the same as with the local display.
9.4
Configuring the measuring device
9.4.1
Basic setup
F
E
BD
SD
A0038238
22
Parameters for the correct operation of the device
BD Blocking distance
SD Safety distance
E
Empty calibration (= zero)
F
Full calibration (= measuring span)
All the functions are grouped together in the
Basic setup (00)
function group. When you
have completed your entries for a function, the next function appears automatically.
Basic setup procedure (example)
Function and selection
•
Tank shape
→ Dome ceiling
•
Medium property
→ Unknown
•
Process cond.
→ Standard
•
Empty calibr.
•
Full calibr.
•
Mapping
Optional functions
•
Safety settings
•
Linearization
•
Extended calibration
•
...