11 Description of device functions
Prosonic Flow 92
70
EHauser
Wall Thickness
Use this function to enter the thickness of the pipe wall.
User input (Metric):
Fixed-point number 0.01...100.00 mm.
User input (US):
Fixed-point number 0.0004...4.000 inch
Factory setting:
4.5 mm
Liner Material
Use this function to select the liner material of the pipe.
Options:
NO LINER – TAR EPOXY – MORTAR – RUBBER – TEFLON – GLASS PYREX –
PVC – OTHER
Note!
If “OTHER” is selected, the sound velocity can be entered in m/s (metric) or
ft/s (US).
Factory setting:
NO LINER
Liner Thickness
Use this function to enter the thickness of the liner.
User input (Metric):
Fixed-point number 0.01...100.00 mm.
User input (US):
Fixed-point number 0.0004...4.000 inch
Liquid
Use this function to select the liquid in the pipe.
Options:
WATER – SEAWATER – DISTILLED WATER – AMMONIA – ALCOHOL – BEN-
ZENE – BROMIDE – ETHANOL – GLYCOL – KEROSENE – MILK – METHA-
NOL – TOLUOL – LUBRICATING OIL – FUEL OIL– PETROL – OTHER
If “OTHER” is selected, the sound velocity can be entered in m/s (metric) or
ft/s (US) and the viscosity in mm
2
/s (metric) or ft
2
/s.
Factory setting:
WATER
Temperature
Use this function to enter the process temperature of the selections liquid.
User input:
Fixed-point number 0...300.00 °C
Factory setting:
20.0 °C
Site Setup
Site Data
Summary of Contents for prosonic flow 92
Page 4: ...Brief operating instructions Prosonic Flow 92 4 Endress Hauser...
Page 12: ...2 Identification Prosonic Flow 92 12 Endress Hauser...
Page 26: ...3 Installation Prosonic Flow 92 26 Endress Hauser...
Page 34: ...4 Wiring Prosonic Flow 92 34 Endress Hauser...
Page 40: ...5 Operation Prosonic Flow 92 40 Endress Hauser...
Page 46: ...7 Maintenance Prosonic Flow 92 46 Endress Hauser...