9 Trouble-shooting
Prosonic Flow 92
50
EHauser
9.2
Error messages
Serious system errors are
always
recognised by the instrument as “Fault messages”,
and are shown on the display. Fault messages immediately affect the inputs and out-
puts.
"
Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer
for repair. The procedures on Page 8 must be carried out before you return a flowmeter
to Hauser. Always enclose a fully completed “Declaration of Contamination”
form. You will find a copy of the form at the back of these Operating Instructions.
!
Note!
Also observe the information on Page 39 and 52.
Error message
Error description
Remedy / spare part
Communication
No data transfer between internal
modules or faulty internal data
transfer.
Spare parts see Page 53
→
contact E+H
Amplifier Fault
Amplifier board defective.
Replace the amplifier board.
Spare parts see Page 53
→
contact E+H
Amplifier Data
Amplifier data invalid.
Verify Site Setup data.
Restart device.
Received Signal
The strength of received signals
is changing.
– Check whether the application
causes interference of the
signal.
– Check for air in the fluid.
– Check for solid content.
– Check for stable flow
conditions.
Init. Run
Initialisation running.
Wait until the procedure is com-
pleted.
Signal Low
Attenuation of acoustic measure-
ment section too high.
– Check to see if coupling fluid
must be renewed.
– It is possible that the fluid
indicates too much attenuation.
– Check the sensor distance
(installation dimensions).
– Reduce the number of traverses
if possible.
Signal High
Received signal is too high.
Reduce burst voltage.
Error Current Output
Range current out exceeded.
Modify the range.
Backup Battery Fail.
The backup battery for setup and
log data failed and needs to be
exchanged.
Spare parts see Page 53
→
contact E+H
Summary of Contents for prosonic flow 92
Page 4: ...Brief operating instructions Prosonic Flow 92 4 Endress Hauser...
Page 12: ...2 Identification Prosonic Flow 92 12 Endress Hauser...
Page 26: ...3 Installation Prosonic Flow 92 26 Endress Hauser...
Page 34: ...4 Wiring Prosonic Flow 92 34 Endress Hauser...
Page 40: ...5 Operation Prosonic Flow 92 40 Endress Hauser...
Page 46: ...7 Maintenance Prosonic Flow 92 46 Endress Hauser...