Troubleshooting
Proline Prosonic Flow 93C FOUNDATION Fieldbus
64
Hauser
043
Device status message (FF):
T-DAT data inconsistent –
Err. No. 043
Local display:
S: C-DATA T-DAT
$
: # 043
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause:
Error in calibration data
Remedy:
1.
Check whether the T-DAT is correctly plugged
into the amplifier board.
2.
Replace the T-DAT if it is defective.
Check that the new, replacement DAT is
compatible with the measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3.
Replace measuring electronic boards if
necessary.
Transducer_Error = Electronics
failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
082
Device status message (FF):
Interruption between sensor
and transmitter CH1 -
Err. No. 082
Local display:
S: SENS. DOWN CH1
$
: # 082
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause:
Connection between sensor channel 1/2 and
transmitter interrupted.
Remedy:
1.
Check the cable connection between the
sensor and the transmitter.
2.
Check that the sensor connector is fully
screwed in.
3.
Check that the correct sensor has been
connected.
4.
Local display:
In the SENSOR TYPE function (6881), check
that the correct sensor has been selected for
Channel 1 or Channel 2.
5.
FOUNDATION Fieldbus:
In the "Sensor Param. - Sensor Type"
parameter, check whether the correct sensor
has been selected for error number 82 or 83.
6.
The sensor may be defective.
Transducer_Error =
Mechanical failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
083
Device status message (FF):
Interruption between sensor
and transmitter CH2 -
Err. No. 083
Local display:
S: SENS. DOWN CH2
$
: # 083
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Transducer_Error =
Mechanical failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
085
Device status message (FF):
Interruption between sensor
and transmitter CH1 -
Err. No. 085
Local display:
S: SENS. UP CH1
$
: # 086
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Cause:
Connection between sensor channel 1/2 and
transmitter interrupted.
Remedy:
1.
Check the cable connection between the
sensor and the transmitter.
2.
Check that the sensor connector is fully
screwed in.
3.
Check that the correct sensor has been
connected.
4.
Check that the correct sensor has been
selected.
– FF: Transducer Block "Transducer_CH1"
(base index: 1400) or Transducer Block
"Transducer_CH2" (base index: 1500)
→
Parameter "Sensor Param. - Sensor Type"
– Onsite display: BASIC FUNCTION
→
SENSOR DATA CH1 or CH2
→
SENSOR
PARAMETERS
→
SENSOR TYPE
5.
The sensor may be defective.
Transducer_Error =
Mechanical failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
086
Device status message (FF):
Interruption between sensor
and transmitter CH2 -
Err. No. 086
Local display:
S: SENS. DOWN CH2
$
: # 086
BLOCK_ERR = Device needs
maintenance now
OUT. QUALITY = BAD
Transducer_Error =
Mechanical failure
OUT. SUBSTATUS = Device
Failure
BLOCK_ERR = Input Failure
(faulty input value from
Transducer Blocks)
No.
Error messages:
FOUNDATION Fieldbus
(FF)*
(local display)
Transducer Block
error messages
"Diagnosis" Transducer
Block
Analog Input Function Block
error messages
Reason for error/rectification
(spare parts
→
ä
71)