Installation
Proline Prosonic Flow 91
14
Hauser
3.3
Preparatory steps prior to installation
Depending on the conditions specific to the measuring point (e.g. clamp-on, number of traverses,
fluid, etc.), a number of preparatory steps have to be taken before actually installing the sensors:
1.
Determination of the values for the necessary installation distances based on the conditions
specific to the measuring point. A number of methods are available for determining the values:
– Local operation of the device
– FieldCare (operating program), connect a notebook to the transmitter
– Applicator (software), online on the Hauser Internet site
2.
Mechanical preparation of the clamp-on retainers for the sensors:
– Premount the strapping bands (DN 50 to 200 / 2 to 8") or (DN 250 to 4000 / 10 to 160")
– Fix the welded bolts
3.4
Determining the necessary installation distances
The installation distance that have to be maintained depend on:
• The type of sensor: W (DN 50 to 4000 / 2 to 160"), W(DN 15 to 65 / ½ to 2½")
• Type of mounting:
– Clamp-on with strapping band or welded bolt
– Insertion version, installation in the pipe
• Number of traverses or single-path/dual-path version
3.4.1
Installation distances for Prosonic Flow W clamp-on
3.4.2
Determining values for installation distances
Perform the following steps to determine the installation distances:
1.
Mount the transmitter.
2.
Connect the power supply.
3.
Switch on the measuring device.
4.
Run the "Sensor Setup menu.
DN 50 to 4000 (2 to 160")
DN 15 to 65 (½ to 2½")
Clamp-on
Strapping band
Clamp-on
Welded bolts
Clamp-on
Strapping band
1 traverse
2 traverses
1 traverse
2 traverses
2 traverses
SENSOR DISTANCE
SENSOR DISTANCE
SENSOR DISTANCE
SENSOR DISTANCE
SENSOR DISTANCE
WIRE LENGTH
POSITION SENSOR
WIRE LENGTH
POSITION SENSOR
–