Proline Promass P 300
42
Hauser
Performance characteristics
Reference operating
conditions
• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the
Applicator
Maximum measured error
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Mass flow and volume flow (liquids)
±0.10 % o.r.
Density (liquids)
Under
reference operating conditions
Standard density
calibration
Wide-range
Density specification
[g/cm³]
[g/cm³]
[g/cm³]
±0.0005
±0.01
±0.002
1)
Valid over the entire temperature and density range
2)
Valid range for special density calibration: 0 to 2 g/cm³, +10 to +80 °C (+50 to +176 °F)
3)
Order code for "Application package", option EE "Special density"
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
DN
Zero point stability
[mm]
[in]
[kg/h]
[lb/min]
8
³⁄₈
0.20
0.007
15
½
0.65
0.024
25
1
1.80
0.066
40
1½
4.50
0.165
50
2
7.0
0.257
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
DN
1:1
1:10
1:20
1:50
1:100
1:500
[mm]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
8
2
000
200
100
40
20
4
15
6
500
650
325
130
65
13
25
18
000
1
800
900
360
180
36