corresponding type test declaration can be ordered with the device (order code for "Additional

approval", option LN "Sensor housing burst pressure, type test").

DN

Sensor housing nominal

pressure

(designed with a safety factor

≥ 4)

Sensor housing burst pressure

[mm]

[in]

[bar]

[psi]

[bar]

[psi]

8

³⁄₈

25

362

190

2

 

755

15

½

25

362

175

2

 

538

25

1

25

362

165

2

 

392

40

25

362

152

2

 

204

50

2

25

362

103

1

 

494

For information on the dimensions: see the "Mechanical construction" section→   57

Flow limit

Select the nominal diameter by optimizing between the required flow range and permissible pressure

loss.

For an overview of the full scale values for the measuring range, see the "Measuring range"

section →   10

• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value

• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal

• A low full scale value must be selected for abrasive media (such as liquids with entrained solids):

flow velocity < 1 m/s (< 3 ft/s).

To calculate the flow limit, use the 

Applicator

 sizing tool →   109

Pressure loss 

To calculate the pressure loss, use the 

Applicator

 sizing tool →   109

System pressure

It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas.

This is prevented by means of a sufficiently high system pressure.
For this reason, the following mounting locations are recommended:

• At the lowest point in a vertical pipe

• Downstream from pumps (no danger of vacuum)

  A0028777

Thermal insulation

In the case of some fluids, it is important to keep the heat radiated from the sensor to the

transmitter to a low level. A wide range of materials can be used for the required insulation.
The following device versions are recommended for versions with thermal insulation:

• Version with extended neck for insulation:

Order code for "Sensor option", option CG with an extended neck length of 105 mm (4.13 in).

• Extended temperature version:

Order code for "Measuring tube material", option TD or TG with an extended neck length of

105 mm (4.13 in).

Summary of Contents for Proline Promass P 300

Page 1: ...IP69 and up to 3 I Os Backlit display with touch control and WLAN access Remote display available Your benefits Highest process quality fully compliant with industry requirements Fewer process measuri...

Page 2: ...let runs 47 Special mounting instructions 47 Environment 49 Ambient temperature range 49 Storage temperature 49 Climate class 49 Degree of protection 49 Vibration resistance 49 Shock resistance 49 Sho...

Page 3: ...ration 106 Special density 106 OPC UA server 106 Accessories 106 Device specific accessories 107 Communication specific accessories 108 Service specific accessories 109 System components 109 Supplemen...

Page 4: ...nects the protectiv earth to the mains supply Outer ground terminal Connects the device to the plant grounding system Communication symbols Symbol Meaning Wireless Local Area Network WLAN Communicatio...

Page 5: ...flow Instead of a constant rotational velocity the sensor uses oscillation In the sensor an oscillation is produced in the measuring tube The Coriolis forces produced at the measuring tube cause a pha...

Page 6: ...vailable as a compact version The transmitter and sensor form a mechanical unit Transmitter Promass 300 A0026708 Device versions and materials Transmitter housing Aluminum coated aluminum AlSi10Mg coa...

Page 7: ...of specific functions to support protective measures on the operator s side These functions can be configured by the user and guarantee greater in operation safety if used correctly An overview of the...

Page 8: ...extra is protected by the network key The WLAN authentication of the network key complies with the IEEE 802 11 standard When the device is delivered the network key is pre defined depending on the dev...

Page 9: ...and encrypted Access via service interface CDI RJ45 The device can be connected to a network via the service interface CDI RJ45 Device specific functions guarantee the secure operation of the device...

Page 10: ...nal Input and output versions 12 External measured values To increase the accuracy of certain measured variables the automation system can continuously write various measured values to the measuring d...

Page 11: ...active 0 4 to 20 mA passive Resolution 1 A Voltage drop Typically 0 6 to 2 V for 3 6 to 22 mA passive Maximum input voltage 30 V passive Open circuit voltage 28 8 V active Possible input variables Pre...

Page 12: ...rder code for Output input 2 021 Not assigned A A A A A A A A A A Current output 0 4 to 20 mA B B B B B B B Current output 0 4 to 20 mA Ex i C C C User configurable input output 1 D D D D D D D Pulse...

Page 13: ...Oscillation damping 0 Signal asymmetry Exciter current 0 The range of options increases if the measuring device has one or more application packages PROFIBUS PA PROFIBUS PA In accordance with EN 5017...

Page 14: ...ssive Load 0 to 700 Resolution 0 38 A Damping Adjustable 0 07 to 999 s Assignable measured variables Mass flow Volume flow Corrected volume flow Density Reference density Temperature Electronic temper...

Page 15: ...low Density Reference density Temperature Electronic temperature Oscillation frequency 0 Oscillation damping 0 Signal asymmetry Exciter current 0 The range of options increases if the measuring device...

Page 16: ...ior Can be set to NO normally open factory setting NC normally closed Maximum switching capacity passive DC 30 V 0 1 A AC 30 V 0 5 A Assignable functions Off On Diagnostic behavior Limit value Mass fl...

Page 17: ...ance with PROFIBUS PA Profile 3 02 EtherNet IP Device diagnostics Device condition can be read out in Input Assembly PROFINET Device diagnostics According to Application Layer protocol for decentraliz...

Page 18: ...us Open Closed Relay output Failure mode Choose from Current status Open Closed Local display Plain text display With information on cause and remedial measures Backlight Red backlighting indicates a...

Page 19: ...7 Option BA Current output 4 to 20 mA HART UN 30 VDC UM 250 VAC Option GA PROFIBUS PA UN 30 VDC UM 250 VAC Option LA PROFIBUS DP UN 30 VDC UM 250 VAC Option MA Modbus RS485 UN 30 VDC UM 250 VAC Option...

Page 20: ...ersion transmitter Order code for Output input 2 Output input 3 Output type Intrinsically safe values or NIFW values Output input 2 Output input 3 24 25 22 23 Option C Current output 4 to 20 mA Ex i U...

Page 21: ...file Earlier models Promass 80 PROFIBUS PA ID No 1528 hex Extended GSD file EH3x1528 gsd Standard GSD file EH3_1528 gsd Promass 83 PROFIBUS PA ID No 152A hex Extended GSD file EH3x152A gsd Standard G...

Page 22: ...rity Auto polarity for automatic correction of crossed TxD and RxD pairs Supported CIP connections Max 3 connections Explicit connections Max 6 connections I O connections Max 6 connections scanner Co...

Page 23: ...ufacturer specific software FieldCare DeviceCare Web browser Device master file GSD can be read out via the integrated Web server of the measuring device Configuration of the device name DIP switches...

Page 24: ...vice Link Capabilities Slot time 4 Min delay between PDU 8 Max response delay 16 System integration Information regarding system integration Operating Instructions 110 Cyclic data transmission Descrip...

Page 25: ...on system integration Operating Instructions 110 Modbus RS485 information Function codes Register information Response time Modbus data map Power supply Terminal assignment Transmitter supply voltage...

Page 26: ...device version ordered 12 Terminal assignment of the remote display and operating module 31 Device plugs available Device plugs may not be used in hazardous areas Device plugs for fieldbus systems Or...

Page 27: ...ctrical connection 2 3 L N P U Connector M12 1 R 1 2 S 1 2 T 1 2 V 1 2 Connector M12 1 Connector M12 1 1 Cannot be combined with an external WLAN antenna order code for Enclosed accessories option P8...

Page 28: ...4 Phoenix part no 1543223 SACC M12MSD 4Q When using the device in a hazardous location use a suitably certified plug Service interface Order code for Accessories mounted option NB Adapter RJ45 M12 ser...

Page 29: ...tion NE21 Current consumption Transmitter Max 400 mA 24 V Max 200 mA 110 V 50 60 Hz 230 V 50 60 Hz Power supply failure Totalizers stop at the last value measured Depending on the device version the c...

Page 30: ...to an M12 connector mounted in the cable entry Therefore the connection to the service interface can be established via an M12 connector without opening the device Network connection via service inter...

Page 31: ...connected at the same time as the existing measuring device display module Only one display or operation unit may be connected to the transmitter at any one time 1 3 81 Vcc 82 Gnd 83 A 84 B 81 Vcc 82...

Page 32: ...ons 39 4 Analog display unit observe maximum load 13 5 Transmitter HART input 2 4 20 mA 4 1 2 3 3 6 5 A0028763 4 Connection example for HART input with a common negative passive 1 Automation system wi...

Page 33: ...x 5 Measuring device 6 Local grounding 7 Bus terminator 8 Potential matching line PROFIBUS DP 2 1 A B 3 4 4 A B A B A0028765 6 Connection example for PROFIBUS DP non hazardous area and Zone 2 Div 2 1...

Page 34: ...Device plug 5 Transmitter EtherNet IP DLR Device Level Ring 1 2 4 5 5 3 A0027544 1 Control system e g PLC 2 Ethernet switch 3 Observe cable specifications 39 4 Connecting cable between the two transmi...

Page 35: ...two transmitters 5 Transmitter FOUNDATION Fieldbus 2 1 3 4 7 8 6 6 6 6 5 6 6 5 A0028768 9 Connection example for FOUNDATION Fieldbus 1 Control system e g PLC 2 Power Conditioner FOUNDATION Fieldbus 3...

Page 36: ...rve cable specifications 3 Distribution box 4 Transmitter Current output 4 20 mA 4 20 mA 2 1 3 A0028758 11 Connection example for 4 20 mA current output active 1 Automation system with current input e...

Page 37: ...nput values 14 Switch output 1 2 3 A0028760 14 Connection example for switch output passive 1 Automation system with switch input e g PLC 2 Power supply 3 Transmitter Observe input values 14 Double pu...

Page 38: ...se output 5 Double pulse output slave phase shifted Relay output 1 2 3 A0028760 17 Connection example for relay output passive 1 Automation system with relay input e g PLC 2 Power supply 3 Transmitter...

Page 39: ...vailable for certain device versions 26 Cable specification Permitted temperature range The installation guidelines that apply in the country of installation must be observed The cables must be suitab...

Page 40: ...of ODVA Organization PROFINET Standard IEC 61156 6 specifies CAT 5 as the minimum category for a cable used for PROFINET CAT 5e and CAT 6 are recommended For more information on planning and installi...

Page 41: ...n 1 Cable length Maximum 300 m 1000 ft see the following table Cross section Cable length for use in Non hazardous area Hazardous area Zone 2 Class I Division 2 Hazardous area Zone 1 Class I Division...

Page 42: ...reference operating conditions Standard density calibration 1 Wide range Density specification 2 3 g cm g cm g cm 0 0005 0 01 0 002 1 Valid over the entire temperature and density range 2 Valid range...

Page 43: ...The outputs have the following base accuracy specifications Current output Accuracy 5 A Pulse frequency output o r of reading Accuracy Max 50 ppm o r over the entire ambient temperature range Repeatab...

Page 44: ...cification special density calibration If the process temperature is outside the valid range 42 the measured error is 0 0001 g cm3 C 0 00005 g cm3 F kg m 3 C 40 0 80 40 80 120 160 200 240 280 320 360...

Page 45: ...1332 BaseAccu A0021339 ZeroPoint BaseAccu 100 A0021333 ZeroPoint MeasValue 100 A0021334 Calculation of the maximum repeatability as a function of the flow rate Flow rate Maximum repeatability in o r Z...

Page 46: ...pipeline Pipe restrictions or the use of an orifice with a smaller cross section than the nominal diameter prevent the sensor running empty while measurement is in progress 1 2 3 4 5 A0028773 20 Insta...

Page 47: ...of the sensor to the fluid properties 1 2 A0028774 21 Orientation of sensor with curved measuring tube 1 Avoid this orientation for fluids with entrained solids Risk of solids accumulating 2 Avoid th...

Page 48: ...ing dimensions must be observed Use mounting clamp with lining between clamp and measuring instrument A B C A0030298 DN A B C mm in mm in mm in mm in 8 298 11 73 33 1 3 28 1 1 15 402 15 83 33 1 3 28 1...

Page 49: ...rm climatic regions You can order a weather protection cover from Endress Hauser 107 Storage temperature 50 to 80 C 58 to 176 F Climate class DIN EN 60068 2 38 test Z AD Degree of protection Measuring...

Page 50: ...terminal wherever possible Details are provided in the Declaration of Conformity Process Medium temperature range Standard version 50 to 150 C 58 to 302 F Order code for Measuring tube mat wetted sur...

Page 51: ...ust the process connection The diagrams show the maximum permissible medium pressure depending on the specific medium temperature Pressure temperature curves with temperature range 151 to 205 C 304 to...

Page 52: ...lange material 1 4404 F316 F316L Flange DIN 11864 2 Form A DN8 40 DN50 80 0 80 160 240 320 400 0 10 20 30 40 0 200 400 600 C F bar psi 200 150 100 50 0 50 50 A0027781 EN 26 With flange material 1 4435...

Page 53: ...sealing materials are used Please take this into account when selecting seals and counterparts as these components can limit the pressure and temperature range Thread DIN 11864 1 Form A DN8 40 DN50 8...

Page 54: ...e electronics and mechanics inside If a measuring tube fails e g due to process characteristics like corrosive or abrasive fluids the fluid will initially be contained by the sensor housing If the sen...

Page 55: ...ted for abrasive media such as liquids with entrained solids flow velocity 1 m s 3 ft s To calculate the flow limit use the Applicator sizing tool 109 Pressure loss To calculate the pressure loss use...

Page 56: ...am Via heating jackets Heating jackets for the sensors can be ordered as accessories from Endress Hauser 108 NOTICE Danger of overheating when heating Ensure that the temperature at the lower end of t...

Page 57: ...G or order code for Measuring tube material option TD TG values 70 mm 3 For version without local display values 30 mm 4 Depends on the process connection in question Order code for Housing option A A...

Page 58: ...15 183 73 110 108 266 374 207 65 142 12 0 3 92 25 183 73 110 121 266 387 207 65 142 17 6 3 92 40 183 73 110 178 297 475 207 65 142 26 0 3 142 50 183 73 110 262 310 572 207 65 142 38 0 3 169 1 Dependi...

Page 59: ...85 4 14 19 0 28 5 580 40 150 110 4 18 21 0 43 1 707 50 165 125 4 18 25 0 54 5 828 Surface roughness flange EN 1092 1 Form B1 DIN 2526 Form C Ra 3 2 to 12 5 m 1 DN 8 with DN 15 flanges as standard Fla...

Page 60: ...e for Process connection option NEW DN mm A mm B mm C mm D mm E mm L mm 8 1 95 70 4 15 16 0 15 336 15 95 70 4 15 16 0 15 440 25 125 90 4 19 17 5 25 580 40 140 105 4 19 20 0 40 707 50 155 120 8 19 27 5...

Page 61: ...2 Form A for pipe according to DIN 11866 series A flange with notch 1 4435 316L Order code for Process connection option KJW DN mm A mm B mm C mm D mm E mm L mm 8 59 0 42 4 9 10 16 00 384 15 59 0 42...

Page 62: ...0 26 00 626 40 82 65 4 9 10 38 00 753 50 94 77 4 9 10 50 00 877 Ra max 0 76 m order code for Measuring tube material option BB TD Ra max 0 38 m electropolished order code for Measuring tube material o...

Page 63: ...t flange sterile orbital for pipe according to DIN 11866 series A flange form R 1 4435 316L Order code for Process connection option BSB DN mm A mm B mm C mm D mm E mm L mm 8 65 45 4 9 10 10 00 384 15...

Page 64: ...or Measuring tube material option BC TG 1 The connection complies with the hygienic clamp dimensions as per ASME BPE Tri Clamp for pipe according to DIN 11866 series C 1 4435 316L Order code for Proce...

Page 65: ...code for Process connection option Clamp in A mm B mm L mm 8 FEB 25 0 9 40 362 15 FED 25 0 15 75 466 25 FEF 1 1 50 4 22 10 606 40 FEH 1 1 50 4 34 80 738 50 FEK 2 1 63 9 47 50 860 Ra max 0 76 m order c...

Page 66: ...nnection option JSA DN mm A mm B mm L mm 8 50 5 22 6 362 15 50 5 22 6 466 25 50 5 22 6 606 40 50 5 35 6 731 50 64 0 48 6 853 Ra max 0 76 m order code for Measuring tube material option BB TD Ra max 0...

Page 67: ...es A 1 4435 316L Order code for Process connection option KLW DN mm A mm B mm L mm 8 34 0 16 05 370 15 34 0 16 05 474 25 50 5 26 05 614 40 64 0 38 05 738 50 77 5 50 05 853 Ra max 0 76 m order code for...

Page 68: ...material option BC TG BBS Quick Connect sterile orbital for pipe according to DIN11866 series B female 1 4435 316L Order code for Process connection option BSJ DN mm A mm B mm L mm 8 25 0 14 00 362 1...

Page 69: ...nection DIN 32676 ISO 2852 L A B A0016543 Length tolerance for dimension L in mm 1 5 2 0 Eccentric clamp DIN 32676 for pipe according to DIN 11866 series A 1 4435 316L Order code for Process connectio...

Page 70: ...r code for Measuring tube material option BB TD Ra max 0 38 m electropolished order code for Measuring tube material option BC TG Additional information on Eccentric clamps Eccentric clamp connection...

Page 71: ...77 5 50 00 860 Ra max 0 76 m order code for Measuring tube material option BB TD Ra max 0 38 m electropolished order code for Measuring tube material option BC TG Additional information on Eccentric...

Page 72: ...e according to DIN 11866 series C clamp form R 1 4435 316L Order code for Process connection option BEA DN mm A mm B mm L mm 8 25 10 00 362 15 25 16 00 466 25 25 26 00 610 50 25 50 00 859 Ra max 0 76...

Page 73: ...rial option BB TD Ra max 0 38 m electropolished order code for Measuring tube material option BC TG Thread DIN 11864 1 Form A for pipe according to DIN 11866 series A 1 4435 316L Order code for Proces...

Page 74: ...pipe according to DIN 11866 series B 1 4435 316L Order code for Process connection option BSD DN mm A mm B mm L mm 8 M26 1 5 10 00 362 15 M30 2 16 00 466 25 M42 2 26 00 606 40 M56 2 38 00 738 50 M68 2...

Page 75: ...864 Ra max 0 76 m order code for Measuring tube material option BB TD Ra max 0 38 m electropolished order code for Measuring tube material option BC TG Accessories Rinse connections L 45 A 1 NPT 35 1...

Page 76: ...te display and operating module DKX001 136 5 35 114 4 49 86 3 39 78 3 07 103 4 06 12 0 47 7 0 0 27 A0028921 45 Engineering unit mm in External WLAN antenna The external WLAN antenna is not suitable fo...

Page 77: ...LAN antenna mounted with cable The external WLAN antenna can be mounted separately from the transmitter if the transmission reception conditions at the transmitter mounting location are poor 70 2 8 15...

Page 78: ...r Measuring tube material option TD TG values 2 76 in 3 For version without local display values 1 18 in 4 Depends on the process connection in question Order code for Housing option A Aluminum coated...

Page 79: ...gland used values up to 1 18 in 2 With order code for Sensor option option CG or order code for Measuring tube material option TD TG values 2 76 in 3 Depends on the process connection in question Flan...

Page 80: ...316L Order code for Process connection option BSB DN in A in B in C in D in E in L in 2 56 1 77 4 0 35 0 39 0 39 15 12 2 95 2 17 4 0 35 0 39 0 63 19 21 1 3 35 2 56 4 0 35 0 47 1 02 24 65 1 3 94 3 15...

Page 81: ...ing tube material option BC TG Tri Clamp for pipe according to DIN 11866 series C 1 4435 316L Order code for Process connection option FFW DN in Clamp in A in B in L in 0 98 0 62 14 25 Ra max 30 in or...

Page 82: ...umo BioConnect clamp connection L X A B X A0016905 50 Detail X Asymmetrical process connection the part shown in blue is provided by the supplier Length tolerance for dimension L in inch 0 06 0 08 Cla...

Page 83: ...pe according to DIN 11866 series C clamp form R 1 4435 316L Order code for Process connection option BEA DN in A in B in L in 0 98 0 39 14 25 0 98 0 63 18 35 1 0 98 1 02 24 02 2 0 98 1 97 43 39 Ra max...

Page 84: ...70 1 91 34 02 Ra max 30 in order code for Measuring tube material option BB TD Ra max 15 in electropolished order code for Measuring tube material option BC TG Accessories Rinse connections L 45 A 1 N...

Page 85: ...te display and operating module DKX001 136 5 35 114 4 49 86 3 39 78 3 07 103 4 06 12 0 47 7 0 0 27 A0028921 53 Engineering unit mm in External WLAN antenna The external WLAN antenna is not suitable fo...

Page 86: ...uding transmitter as per order code for Housing option A Aluminum coated Different values due to different transmitter versions Transmitter version for the hazardous area Order code for Housing option...

Page 87: ...es are suitable for hazardous and non hazardous areas Cable entry cable gland Material Cable gland M20 1 5 Plastic nickel plated brass Adapter for cable entry with internal thread G Nickel plated bras...

Page 88: ...B2220 Stainless steel 1 4404 316 316L All other process connections Stainless steel 1 4435 BN2 316L Available process connections 89 Seals Welded process connections without internal seals Accessorie...

Page 89: ...Neumo BioConnect clamp DIN 11866 series A clamp form R Thread DIN 11851 thread DIN 11866 series A SMS 1145 thread ISO 2853 thread ISO 2037 DIN 11864 1 Form A thread DIN 11866 series A BBS thread ster...

Page 90: ...lements 4 line illuminated graphic display White background lighting switches to red in event of device errors Format for displaying measured variables and status variables can be individually configu...

Page 91: ...ting module DKX001 depends on the choice of transmitter housing material Transmitter housing Remote display and operating module Order code for Housing Material Material Option A Aluminum coated AlSi1...

Page 92: ...4 5 7 6 9 8 A0028746 60 Options for remote operation via HART protocol passive 1 Control system e g PLC 2 Transmitter power supply unit e g RN221N with communication resistor 3 Connection for Commubo...

Page 93: ...FF HSE FF H1 6 FOUNDATION Fieldbus FF H1 network 7 Power supply FF H1 network 8 T box 9 Measuring device Via PROFIBUS DP network This communication interface is available in device versions with PROFI...

Page 94: ...This communication interface is available in device versions with a Modbus RS485 output 1 2 3 A0029437 64 Options for remote operation via Modbus RS485 protocol active 1 Control system e g PLC 2 Comp...

Page 95: ...ring device Ring topology The device is integrated via the terminal connection for signal transmission output 1 and the service interface CDI RJ45 4 2 1 3 5 5 5 A0033725 66 Options for remote operatio...

Page 96: ...DTM CDI Communication TCP IP 3 Switch e g Scalance X204 Siemens 4 Measuring device Ring topology This communication interface is available in device versions with PROFINET 3 1 2 4 4 4 A0033719 68 Opti...

Page 97: ...Communication TCP IP or Modbus DTM 2 Standard Ethernet connecting cable with RJ45 connector 3 Service interface CDI RJ45 of the measuring device with access to the integrated Web server Via WLAN inter...

Page 98: ...the device can be integrated into an Ethernet network via the service interface CDI RJ45 and WLAN and communicate with OPC UA clients If the device is used in this way IT security must be considered...

Page 99: ...gement Solutions AMS by Emerson www emersonprocess com FieldCommunicator 375 475 by Emerson www emersonprocess com Field Device Manager FDM by Honeywell www honeywellprocess com FieldMate by Yokogawa...

Page 100: ...r data nominal diameter etc Serial number Calibration data Device configuration e g SW options fixed I O or multi I O Storage location Fixed on the user interface board in the connection compartment A...

Page 101: ...to 250 measured values via each of the 4 memory channels Export the measured value log via a variety of interfaces and operating tools e g FieldCare DeviceCare or web server Certificates and approval...

Page 102: ...Db CCSAUS Currently the following versions for use in hazardous areas are available IS Ex i and XP Ex d Class I III III Division 1 Groups A G Class I III III Division 1 Groups C G NI Ex nA Class I Di...

Page 103: ...ieldbus certification FOUNDATION Fieldbus interface The measuring device is certified and registered by the FieldComm Group The measuring system meets all the requirements of the following specificati...

Page 104: ...s and certificates Pressure test internal procedure inspection certificate EN10204 3 1 material certificate wetted parts and sensor housing PMI test XRF internal procedure wetted parts test report EN1...

Page 105: ...der directly in the Endress Hauser Online Shop Application packages Many different application packages are available to enhance the functionality of the device Such packages might be needed to addres...

Page 106: ...and outputting of fluid concentrations The measured density is converted to the concentration of a substance of a binary mixture using the Concentration application package Choice of predefined fluid...

Page 107: ...the separate product structure DKX001 Mounting bracket for DKX001 Ordered directly with the DKX001 Order code for Enclosed accessories option RA Mounting bracket 1 2 pipe If ordered subsequently orde...

Page 108: ...on TI00025S Operating Instructions BA00053S Fieldgate FXA520 Gateway for the remote diagnostics and remote configuration of connected HART measuring devices via a Web browser Technical Information TI0...

Page 109: ...clemanagement FieldCare FDT based plant asset management tool from Endress Hauser It can configure all smart field units in your system and helps you manage them By using the status information it is...

Page 110: ...roline 300 KA01309D KA01229D KA01227D KA01386D KA01311D KA01339D KA01341D Operating Instructions Measuring device Documentation code HART FOUNDATION Fieldbus PROFIBUS PA PROFIBUS DP Modbus RS485 Ether...

Page 111: ...ntents Documentation code HART FOUNDATION Fieldbus PROFIBUS PA PROFIBUS DP Modbus RS485 PROFINET EtherNet IP Web server SD01662D SD01665D SD01664D SD02226D SD01663D SD01969D SD01968D Heartbeat Technol...

Page 112: ...Proline Promass P 300 PROFINET Registered trademark of the PROFIBUS User Organization Karlsruhe Germany TRI CLAMP Registered trademark of Ladish Co Inc Kenosha USA www addresses endress com...

Reviews: