Diagnostics and troubleshooting
Proline Promass E 300 HART
162
Hauser
12.11.1 Function scope of the "Device reset" parameter
Options
Description
Cancel
No action is executed and the user exits the parameter.
To delivery settings
Every parameter for which a customer-specific default setting was ordered is reset
to this customer-specific value. All other parameters are reset to the factory
setting.
Restart device
The restart resets every parameter whose data are in the volatile memory (RAM)
to the factory setting (e.g. measured value data). The device configuration remains
unchanged.
Restore S-DAT backup
Restore the data that are saved on the S-DAT. The data record is restored from the
electronics memory to the S-DAT.
This option is displayed only in an alarm condition.
12.12 Device information
The
Device information
submenu contains all parameters that display different
information for device identification.
Navigation
"Diagnostics" menu → Device information
‣
Device information
Device tag
Serial number
Firmware version
Device name
Manufacturer
Order code
Extended order code 1
Extended order code 2
Extended order code 3
ENP version
Device revision
Device ID