Diagnostics and troubleshooting
Proline Promass E 300 HART
154
Hauser
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
140
Sensor signal
asymmetrical
1. Check or replace sensor electronic
module (ISEM)
2. If available: Check connection cable
between sensor and transmitter
3. Replace sensor
S
Alarm
144
Measurement error too
high
1. Check or change sensor
2. Check process conditions
F
Alarm
Diagnostic of electronic
201
Device failure
1. Restart device
2. Contact service
F
Alarm
242
Software incompatible
1. Check software
2. Flash or change main electronics
module
F
Alarm
252
Modules incompatible
1. Check electronic modules
2. Check if correct modules are
available (e.g. NEx, Ex)
3. Replace electronic modules
F
Alarm
252
Modules incompatible
1. Check if correct electronic modul is
plugged
2. Replace electronic module
F
Alarm
262
Sensor electronic
connection faulty
1. Check or replace connection cable
between sensor electronic module
(ISEM) and main electronics
2. Check or replace ISEM or main
electronics
F
Alarm
270
Main electronic failure
Change main electronic module
F
Alarm
271
Main electronic failure
1. Restart device
2. Change main electronic module
F
Alarm
272
Main electronic failure
1. Restart device
2. Contact service
F
Alarm
273
Main electronic failure
Change electronic
F
Alarm
275
I/O module 1 to n
defective
Change I/O module
F
Alarm
276
I/O module 1 to n faulty 1. Restart device
2. Change I/O module
F
Alarm
283
Memory content
1. Reset device
2. Contact service
F
Alarm
302
Device verification active Device verification active, please wait.
C
Warning
303
I/O 1 to n configuration
changed
1. Apply I/O module configuration
(parameter 'Apply I/O
configuration')
2. Afterwards reload device
description and check wiring
M
Warning
311
Electronic failure
1. Do not reset device
2. Contact service
M
Warning
332
Writing in HistoROM
backup failed
Replace user interface board
Ex d/XP: replace transmitter
F
Alarm
361
I/O module 1 to n faulty 1. Restart device
2. Check electronic modules
3. Change I/O Modul or main
electronics
F
Alarm