Proline Promass E 300 HART
Diagnostics and troubleshooting
Hauser
157
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
831
Sensor temperature too
low
Increase ambient temp. around the
sensor housing
S
Warning
832
Electronic temperature
too high
Reduce ambient temperature
S
Warning
833
Electronic temperature
too low
Increase ambient temperature
S
Warning
834
Process temperature too
high
Reduce process temperature
S
Warning
835
Process temperature too
low
Increase process temperature
S
Warning
842
Process limit
Low flow cut off active!
1. Check low flow cut off configuration
S
Warning
862
Partly filled pipe
1. Check for gas in process
2. Adjust detection limits
S
Warning
882
Input signal
1. Check input configuration
2. Check external device or process
conditions
F
Alarm
910
Tubes not oscillating
1. Check electronic
2. Inspect sensor
F
Alarm
912
Medium inhomogeneous 1. Check process cond.
2. Increase system pressure
S
Warning
913
Medium unsuitable
1. Check process conditions
2. Check electronic modules or sensor
S
Warning
941
API temperature out of
specification
1. Check process temperature with
selected API commodity group
2. Check API related parameters
S
Warning
942
API density out of
specification
1. Check process density with selected
API commodity group
2. Check API related parameters
S
Warning
943
API pressure out of
specification
1. Check process pressure with
selected API commodity group
2. Check API related parameters
S
Warning
944
Monitoring failed
Check process conditions for Heartbeat
Monitoring
S
Warning
948
Oscillation damping too
high
Check process conditions
S
Warning
1)
Diagnostic behavior can be changed.
12.8 Pending diagnostic events
The
Diagnostics
menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool → 152
• Via "DeviceCare" operating tool → 152
Other pending diagnostic events can be displayed in the
Diagnostic list
submenu
Summary of Contents for Proline Promass E 300 HART
Page 211: ......