Proline Promag
Installation
Hauser
17
Thermal insulation Promag P 300/500
Pipes generally have to be insulated if they carry very hot fluids to avoid energy losses and
prevent accidental contact with pipes at temperatures that could cause injury. Guidelines
regulating the insulation of pipes have to be taken into account.
L
WARNING
Electronics overheating on account of thermal insulation!
‣
The housing support dissipates heat and its entire surface area must remain uncovered.
Make sure that the sensor insulation does not extend past the top of the two sensor half-
shells.
max.
A0031216
Vibrations
L
A0029004
6
Measures to avoid device vibrations (L > 10 m (33 ft))