Installation
Proline Promag L 800
22
Hauser
Mounting the seals
"
Caution!
Risk of short circuit! Do not use electrically conductive sealing compounds such as graphite! An
electrically conductive layer could form on the inside of the measuring tube and short-circuit the
measuring signal.
Comply with the following instructions when installing seals:
• For hard rubber lining: additional seals are always required.
• For polyurethane lining: generally additional seals are not required.
• For PTFE lining: generally additional seals are not required.
• For DIN flanges: only use seals according to EN 1514-1.
• Make sure that the seals do not protrude into the piping cross-section.
Mounting the ground cable
• Comply with the information on potential equalization and detailed mounting instructions for
• If necessary, you can order special ground cables for potential equalization as an accessory.