Proline Promag H 300
80
Hauser
Spacer
1.4435 (F316L)
Fitted electrodes
• 2 measuring electrodes for signal detection
• 1 empty pipe detection electrode for empty pipe detection/temperature measurement (only DN 15
to 150 (½ to 6"))
Process connections
With O-ring seal
• Welding nipple (DIN EN ISO 1127, ODT/SMS, ISO 2037)
• Flange (EN (DIN), ASME, JIS)
• Flange from PVDF (EN (DIN), ASME, JIS)
• External thread
• Internal thread
• Hose connection
• PVC adhesive sleeve
With aseptic molded seal:
• Coupling (DIN 11851, DIN 11864-1, ISO 2853, SMS 1145)
• Flange DIN 11864-2
For information on the different materials used in the process connections → 79
Surface roughness
Stainless steel electrodes, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); platinum; tantalum:
≤ 0.3 to 0.5 µm (11.8 to 19.7 µin)
(All data relate to parts in contact with fluid)
Liner with PFA:
≤ 0.4 µm (15.7 µin)
(All data relate to parts in contact with fluid)
Stainless steel process connections:
• With O-ring seal: ≤ 1.6 µm (63 µin)
• With aseptic seal: ≤ 0.8 µm (31.5 µin)
Optional: ≤ 0.38 µm (15 µin)
(All data relate to parts in contact with fluid)
Operability
Operating concept
Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Fast and safe commissioning
• Guided menus ("Make-it-run" wizards) for applications
• Menu guidance with brief explanations of the individual parameter functions
• Device access via Web server
• Optional: WLAN access to device via mobile handheld terminal
Reliable operation
• Operation in local language → 81
• Uniform operating philosophy applied to device and operating tools
• If replacing electronic modules, transfer the device configuration via the integrated memory
(integrated HistoROM) which contains the process and measuring device data and the event
logbook. No need to reconfigure.
Efficient diagnostics increase measurement availability
• Troubleshooting measures can be called up via the device and in the operating tools
• Diverse simulation options, logbook for events that occur and optional line recorder functions