Operation
Promag 50
68
Hauser
5.4.6
Device status and error messages
You can read the extended device status, in this case, current error messages, via Command "48".
The command delivers information which is partly coded in bits (see table below).
!
Note!
• You can find a detailed explanation of the device status and error messages and their elimination
• Bits and bytes not listed are not assigned.
Byte
Bit
Error No.
Short error description
0
0
001
Serious device error
1
011
Measuring amplifier has faulty EEPROM
2
012
Error when accessing data of the measuring amplifier EEPROM
1
1
031
S-DAT: defective or missing
2
032
S-DAT: Error accessing saved values
5
051
I/O and the amplifier are not compatible.
3
3
111
Totalizer checksum error
4
121
I/O board and amplifier not compatible.
4
3
251
Internal communication fault on the amplifier board.
4
261
No data reception between amplifier and I/O board
5
0
321
Coil current of the sensor is outside the tolerance.
7
339
Flow buffer:
The temporarily buffered flow portions (measuring mode for pulsating flow) could not be
cleared or output within 60 seconds.
6
0
340
1
341
2
342
3
343
Frequency buffer:
The temporarily buffered flow portions (measuring mode for pulsating flow) could not be
cleared or output within 60 seconds.
4
344
5
345
6
346
7
347
Pulse buffer:
The temporarily buffered flow portions (measuring mode for pulsating flow) could not be
cleared or output within 60 seconds.
7
0
348
1
349
2
350
3
351
Current output:
Flow is out of range.
4
352
5
353
6
354
7
355
Frequency output:
Flow is out of range.
8
0
356
1
357
2
358