Vibrations
Secure the piping and the sensor if vibration is severe.
Caution:
It is advisable to install the sensor and transmitter separately if vibration is exces-
sively severe. The sensor is shock and vibration resistant for acceleration up to 2 g
in accordance with IEC 68-2-6.
Inlet and outlet runs
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows,
etc. Compliance with the following requirements for the inlet and outlet runs is
necessary in order to ensure measuring accuracy:
• Inlet run
≥
5 x pipe diameter
• Outlet run
≥
2 x pipe diameter
Adapters
With the help of the appropriate adapters (reducers and expanders) the sensor can
be mounted in a pipeline of a larger diameter. For slowly flowing fluids, the resulting
higher velocity increases the measuring accuracy.
> 30 ft
(10 m)
5 x D