Proline Promag 53 PROFIBUS DP/PA
Operation
Hauser
75
5.2.5
Controlling the batching processes using the local display
Filling processes can be carried out directly by means of the local display with the aid of the optional
“(Batching)” software package (F-CHIP, accessories
Page 132). Therefore, the device can be
fully deployed in the field as a “batch controller.”
Procedure:
1.
Configure all the required batching functions and assign the lower display info line (=
BATCHING KEYS) using the “Batch” Quick Setup menu (
matrix (
The following “softkeys” then appear on the bottom line of the local display
– START = left display key (
S
)
– PRESET = middle display key (
O
)
– MATRIX = right display key (
F
)
2.
Press the “PRESET (
O
)” key. Various batching process functions requiring configuration will
now appear on the display:
3.
After exiting the PRESET menu, you can now start the batching process by pressing “START
(
S
)”. New softkeys (STOP/HOLD or GO ON) now appear on the display. You can use these
to interrupt, continue or stop the batching process at any time.
STOP
(
S
)
→
Stops batching process
HOLD
(
O
)
→
Interrupts batching process (softkey changes to “GO ON”)
GO ON
(
O
)
→
Continues batching process (softkey changes to “HOLD”)
After the batch quantity is reached, the “START” or “PRESET” softkeys reappear on the
display.
a0004386
Fig. 46:
Controlling batching processes using the local display (softkeys)
“PRESET”
→
Initial settings for the batching process
No.
Function
Settings
7200
BATCH SELECTOR
OS
→
Selection of the batching fluid (BATCH #1 to 6)
7203
BATCHING QUANTITY
If the “ACCESS CUSTOMER” option was selected for the “PRESET
batch quantity” prompt in the “Batching” Quick Setup, the batching
quantity can be altered via the local display. If the “LOCKED” option
was selected, the batching quantity can only be read and cannot be
altered until the private code has been entered.
7265
RESET TOTAL QUANTITY/
TOTALIZER
Resets the batching quantity counter or the total batching quantity to
“0”.
STOP
GO ON
MATRIX
+2.5 l
0.0 l
+2.5 l
0.0 l
0.0 l
+
-
+2.5 l
0.0 l
START
PRESET
MATRIX
STOP
HOLD
MATRIX
E