Technical Information

Oxymax W COS71

Potentiostatic amperometric three-electrode sensor
Long-term stable sensor for trace measurement

Application

Continuous measurement of the dissolved oxygen 
concentration plays an important role in monitoring boiler feed 
water:
• Trace quantities

Oxygen may only be present in trace quantities. Excessive 
concentrations of oxygen can result incorrosion damage in the 
circuit.

• Thermal resp. chemical degassing

Boiler feed water is thermally and chemically degassed to 
achieve oxygen-free water. This state must be maintained in 
service under process conditions.

• Laboratory measurement vs. continuous process monitoring

Spot laboratory measurements to measure oxygen involve a 
risk of error due to sampling errors. Continuous measurement 
directly in the process eliminates this risk and provides 
reliable information about the state of the medium at any 
time.

Your benefits

• Maximum measurement accuracy: 

– Longterm-stable measurement thanks to potentiostatic 

amperometric three-electrode system

– Long maintenance intervals
– Intelligent sensor self monitoring

• Calibration to any transmitter or location and subsequent 

installation at measuring point (in digital mode with Liquisys 
M COM 223/253) as calibration data is stored in the sensor

• Membrane covered sensor, i.e.: 

– high O

2

 selectivity

– Minimum maintenance effort
– Minimum calibration effort thanks to simple calibration in 

air

Summary of Contents for Oxymax W COS71

Page 1: ...itions Laboratory measurement vs continuous process monitoring Spot laboratory measurements to measure oxygen involve a risk of error due to sampling errors Continuous measurement directly in the proc...

Page 2: ...er bromide or silver chloride coating on the anode uses up the bromide or chloride ions dissolved in the electrolyte In the case of conventional membrane covered sensors working with the two electrode...

Page 3: ...el mounting device Remove the SXP connector transmitter side from the cable Refer to the following table for the cable assignment and the assigned terminals for Liquisys M COM223 WX WS Please note tha...

Page 4: ...g l 0 001 ppm Maximum measured error 1 of measured value2 Long term drift with permanent polarization 1 per month Influence of medium pressure Pressure compensation not necessary Polarization time 60...

Page 5: ...ensors You will find them in the Operating Instructions for the assembly used Environment Ambient temperature range 5 to 50 C 20 to 120 F Storage temperature filled with electrolyte 5 to 50 C 20 to 12...

Page 6: ...plug in connection 0 3 kg 0 66 lbs Material Process connection G1 Maximum cable length max 100 m 328 ft including cable extension Membrane thickness approx 25 m Electrolyte Alkaline electrolyte a0004...

Page 7: ...Flow assembly COA260 for plate or wall mounting for oxygen trace measurements ordering acc to product structure Technical Information TI310C 07 en Zero solution 3 units to produce 3 x 1 liter oxygen...

Page 8: ...le Ordering acc to product structure see Technical Information TI 199C 07 en Instruments International Endress Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel 41 61 71...