Oxymax COS61D
Calibration and adjustment
Hauser
23
Altitude
[m (ft)]
K
Altitude
[m (ft)]
K
Altitude
[m (ft)]
K
Altitude
[m (ft)]
K
100 (330)
0.988
650 (2130)
0.927
1150 (3770) 0.874
1650 (5410) 0.825
150 (490)
0.982
700 (2300)
0.922
1200 (3940) 0.869
1700 (5580) 0.820
200 (660)
0.977
750 (2460)
0.916
1250 (4100) 0.864
1750 (5740) 0.815
250 (820)
0.971
800 (2620)
0.911
1300 (4270) 0.859
1800 (5910) 0.810
300 (980)
0.966
850 (2790)
0.905
1350 (4430) 0.854
1850 (6070) 0.805
350 (1150)
0.960
900 (2950)
0.900
1400 (4600) 0.849
1900 (6230) 0.801
400 (1320)
0.954
950 (3120)
0.895
1450 (4760) 0.844
1950 (6400) 0.796
450 (1480)
0.949
1000 (3300) 0.890
1500 (4920) 0.839
2000 (6560) 0.792
500 (1650)
0.943
3. Calculate factor
L
:
Relative air pressure at calibration
L=
-----------------------------------------------
1013 hPa
4. Determine the
M
factor:
•
M
= 1.02 (for
Air 100% rh
calibration method)
•
M
= 1.00 (for
H2O air-saturated
calibration method)
5. Calculate calibration value
C
:
C =
S
.
K
.
L
.
M
Example
• Air calibration at 18 °C (64 °F), altitude 500 m (1650 ft) above sea level, current air
pressure 1009 hPa
• S = 9.45 mg/l, K = 0.943, L = 0.996, M=1.00
• Calibration value C = 8.88 mg/l.
Factor K in the table is not required if the measuring device returns the absolute air
pressure L
abs
(air pressure depending on altitude) as the measured value. The formula
for calculation is then: C = S
.
L
abs
.