Technical data
LPGmass Modbus RS485
74
Hauser
16
Technical data
16.1 Application
The measuring device is suitable for flow measurement of liquids and gases only.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
To ensure that the device remains in proper operating condition for its service life, use the
measuring device only for media against which the process-wetted materials are
sufficiently resistant.
16.2 Function and system design
Measuring principle
Mass flow measurement based on the Coriolis measuring principle
Measuring system
The device consists of a transmitter and a sensor. The Safety Barrier Promass 100 is part
of the scope of supply and must be implemented to operate the device.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
For information on the structure of the device → 11
16.3 Input
Measured variable
Direct measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Corrected volume
Measuring range
Measuring ranges (not custody transfer mode)
DN
Measuring range full scale values
min(F)
to
max(F)
[mm]
[in]
[kg/h]
[lb/min]
8
³⁄₈
0 to 2
000
0 to 73.5
15
½
0 to 6
500
0 to 238
25
1
0 to 18
000
0 to 660
40
1½
0 to 45
000
0 to 1
650
50
2
0 to 70
000
0 to 2
570
The values of the corresponding custody transfer certificate apply in custody transfer
mode.