Environmental
conditions,
continued
Ingress protection IP 65/NEMA4 with closed cover
Vibration security IEC 654-3, v<3mm/s, 1<f<150 Hz
EMC/Immunity
RF protection
To EN 55011 Group 1, Class B
Interference safety
-
ESD
To EN 61000-4-2, Level 3, 6/8 kV
-
Electromagnetic
fields
To EN 61000-4-3, Level 3,103 V/m
-
Burst (supply
circuit)
To EN 61000-4-4, Level 3, 1 kV / 2 kV
-
Burst (signal
circuit)
To EN 61000-4-4, Level 3, 1 kV
-
Surge,
HF discharge
To EN 61000-4-6, 10 V
additional measurement accuracy
≤
0.5%
-
Normal mode
noise rejection
26 dB at input range/10, f = 50/60 Hz, not on
resistance measurement
Housing
construction
Dimensions
W: 100 mm / H: 100 mm / D: 60 mm
Weight
approx. 0.520 kg
Material
Housing: Aluminium die cast, surface galvanised
Wall/stand pipe adapter: 1.4301
Strap: 1.4301
Electrical
connections
Two wire connection (three wire on Pt 100).
Connection access using 2 PG 9 cable glands
(optionally 1 x
½
“ NPT thread instead of 1 x PG9).
Termination on 2,5 mm
2
terminals, 1,5 mm
2
core with
ferrule.
Display and
operating level
Display
LC-Display, 7 segment, prefix, decimal point, limit
symbol, battery status symbol
Operating level
PC software package for setting up , transmission
and display of measured data. Software will run
under Windows 3.11/9x/NT/2000.
Power supply
Power supply
Lithium battery 3.6 Volt Type AA, optionally Type C or
external power supply 7 ... 30 VDC, approx. 5 mA
Battery life cycle
Type AA (2,1 Ah)Type C (7,2 Ah)
Monthly readout:
min. 2 Years
min. 5 Years
Continuous readout: min. 1 month
min. 2 month
Certificates
CE
89/336/EWG guide lines
Technical alterations reserved
12. Technical data