10. Faults and cures
Fault messages are indicated on the display as „E”
followed by a four digit number.
Each number indicates a particular fault condition:
-
„0” = no fault
-
„1” = fault is present
Display
Cause
Effect
Cause
Exxx1
There are no valid
values in the
memory (RAM).
This condition
could be caused by
changing the
battery.
Loss of stored
measured values,
Counter values as
well as date and
time.
Set the unit up
again using the PC
software.
Exx1x
No valid operating
parameters in the
EEPROM.
Loss of operating
settings.
Set up the unit
again using the PC
software .
Ex1xx
No valid calibration
values in the
EEPROM.
Unit not ready for
operation.
Please return the
unit for repair.
E1xxx
CPU error, This
means fault in the
electronics.
Unit defective.
Please return the
unit for repair.