The Micropilot is commissioned by entering an empty distance E (=zero), a full distance
F (=span), and an application parameter. The application parameter automatically
adapts the instrument to the measuring conditions. The data points “E” and “F”
correspond with 4mA and 20mA for instruments with current output. They correspond
with 0 % and 100 % for digital outputs and the display module.
A linearization with max. 32 points, based on a table entered either manually or semi-
automatically, can be activated locally or remotely. This function provides a
measurement in engineering units and a linear output signal for spheres, horizontal
cylindrical tanks, and vessels with conical outlet.
11.4.2 Equipment architecture
Stand-alone
The Micropilot S can be used for measurement in a stilling well as well as in free space.
The different instrument versions are applied as follows:
•
The Micropilot S FMR 532 with planar antenna is the preferred device in stilling wells
≥
150 mm.
•
The Micropilot S FMR 530 with horn antenna can be used in stilling wells with a
diameter <150 mm. However, for these diameters no custody transfer approval is
available.
•
The Micropilot S FMR 533 with parabolic antenna is preferred for free-space
measurements. The Micropilot S FMR 530 with horn antenna can be used as an
alternative for small nozzle diameters .
•
The Micropilot S FMR 531 with rod antenna (PTFE) should be used for measurements
of highly aggressive media (e.g. sulphur).
•
The instruments are equipped with a passive 4…20 mA output with HART protocol.
The complete measuring system consists of:
E HAUSER
E
+
–
%
1
# % &
Copy
G H I
P Q R S
, ( ) ‘
A B C
Paste
Page
On
Page
Up
Delete
Bksp
Insert
J K L
T U V
_ < >
D E F
Hot Key
+ Hot Key
M N O
W X Y Z
+ * /
4
7
.
2
5
8
0
375
FIELD COMMUNICATOR
3
6
9
-
9
6
DELTABAR: * * * * * * * *
ONLINE
1 QUICK SETUP
2 OPERATING MENU
4 SV
0 °C
3 PV
352 mbar
HELP
SAVE
dsdmdm
df das.
asdas fa
asas la.
1
# % &
Copy
G H I
P Q R S
, ( ) ‘
A B C
Paste
Page
On
Page
Up
Delete
Bksp
Insert
J K L
T U V
_ < >
D E F
Hot Key
+ Hot Key
M N O
W X Y Z
+ * /
4
7
.
2
5
8
0
375
FIELD COMMUNICATOR
3
6
9
-
9
6
DELTABAR: * * * * * * * *
ONLINE
1 QUICK SETUP
2 OPERATING MENU
4 SV
0 °C
3 PV
352 mbar
HELP
SAVE
dsdmdm
df das.
asdas fa
asas la.
FXA 191
oder
DXR 375
SPS
Commubox
FXA 191
Service-Adapter
FXA 193
- COMMUWIN II
- ToF Tool
- ToF Tool
Anzeige- und
Bedienmodul
VU 331
HART
HAR
T
2x Messumformer-
speisegerät
RN 221 N
HART-Handbediengerät
DXR 375
L
00-F
M
R53x
x
x
-14-00-06
-e
n
-0
0
2