Display "custody mode" (0A9)
This indicates the instrument calibration mode. The calibration mode (active) can be set
using the hardware security lock on the electronics (see Page 36).
Selection:
•
inactive
•
active pos.
•
active neg.
active pos.
The custody mode (instrument is lead-sealed and accurate to the nearest mm) is active
and is held.
active neg.
Custody mode (instrument is lead-sealed and accurate to the nearest mm) is activated
and not held, e.g. because the signal-to-noise ratio is less than 10 dB (refer to
function "
echo quality
"
(056)
in the function group "
extended calibr.
"
(05)
).
"
Caution!
After entering all the values and completing mounting and aligning work, enter the Reset
Code "
555
" in the function "
reset
"
(0A3)
to reset the instrument history for auto-
correction.
⇒