Micropilot FMR10
Operating menu
Hauser
53
Selection
• Slow <10 cm (0,4 in)/min
• Standard <1 m (40 in)/min
• Fast >1 m (40 in)/min
• No filter / test
Factory setting
Standard <1 m (40 in)/min
First Echo sensitivity
Navigation
Setup → Advanced setup → First Echo sensitivity
Description
This parameter describes the band for First Echo evaluation. Is measured / calculated
down from the peak of the current level echo. Options to select from: - Low The band for
the first echo evaluation is very narrow. The evaluation stays longer at the found echo
respectively does not jump to the next Echo or distortion signal. - Medium The band for
the first echo evaluation has an average width. - High The band for the first echo
evaluation is broad. The evaluation jumps earlier to the next echo or distortion signal.
Selection
• Low
• Medium
• High
Factory setting
Medium
Output mode
Navigation
Setup → Advanced setup → Output mode
Description
Select output mode between: Ullage = The remaining space in the tank or silo is indicated.
or Level linearized = The level is indicated (more precisely: the linearized value if a
linearization has been activated).
Selection
• Ullage
• Level linearized
Factory setting
Level linearized
Blocking distance
Navigation
Setup → Advanced setup → Blocking distance
Description
Specify blocking distance (BD). No signals are evaluated within the blocking distance.
Therefore, BD can be used to suppress interference signals in the vicinity of the antenna.
Note: The measuring range should not overlap with the blocking distance.
User entry
0.0 to 12 m