Liquistation CSF34
Operation
Hauser
91
Menu/Setup/Sampling programs/Setup program/New/Basic
Function
Options
Info
Multiplier
(for version with vacuum pump or
sampling assembly)
1 to 10
Factory setting
1
You can change the sampling volume using
the multiplier. For example, if the dosing
volume is set to 200 ml, set the multiplier to
2 for a sampling volume of 400 ml. 2
samples are taken in succession when
sampling.
Bottle change mode
Selection
• Number of samples
• Time
• External signal
Factory setting
External signal
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
If is selected:
Bottle change mode
Number of samples
Samples per bottle
1 ... 9999
Factory setting
1
Set the number of samples.
Having selected
Bottle change mode
Time
Time interval
00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
00
‐
01:00 DD
‐
HH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Multiple bottles
0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles
"Simultaneous" transfer of two samples to
separate bottles.
Bottle synchronization
Selection
• None
• 1. bottle change time
• 1. Time of
bottle number
Factory setting
None
None
The sampling and bottle change times are
not synchronized.
1. bottle change time
Sampling starts with the first bottle. Set the
synchronization time.
1. Time of bottle number
Each bottle is assigned to a specific fill time.
Set the synchronization time and the
weekday.
Having selected
Bottle change mode
External signal
Bottle chg. sig. input
Selection
• No bottle change input
configured
•
Binary input
Sx
Factory setting
No bottle change input
configured
The bottle change input can be configured
under
Inputs
.
A binary input must be configured for this
function. The sampling input can be
configured under "Inputs".
Multiple bottles
0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles
"Simultaneous" transfer of two samples to
separate bottles.
Start condition
Selection
• Immediate
• Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.