Operation
Liquistation CSF34
104
Hauser
Settings for the Advanced program
Menu/Setup/Sampling programs
Function
Options
Info
Setup program
New
A list of all the programs created is
displayed. For this reason, it is often helpful
to add an "S" for Standard in the program
name.
Advanced
Program name
Free text
Use a distinct name for your sampling
program. The program name can be up to 16
characters long.
Bottle configuration
Choice of all possible bottle
combinations
The ordered bottle configuration is preset or
the configuration selected in the setup is
displayed.
Bottle volume
0 to 100000 ml
Factory setting
• 30000 ml
• 20000 ml
Set the bottle volume. The preset value
depends on the bottle configuration
configured. The bottle volume is always 30 l
for individual containers. The preset value
depends on the bottle configuration. The
bottle volume is always 20 l for individual
containers.
In the case of asymmetric distribution, e.g. 6
x 3 l + 2 x 13 l, you can set the bottle volume
on the left and right in the menu items that
follow.
Start condition
Selection
• Immediate
• Date/time
• Volume
• External start
• External duration
•
Fieldbus
(optional)
Factory setting
Immediate
Immediate
The sampling program is started
immediately.
Date/time
The sampling program is started at a specific
time that can be configured.
Volume
The sampling program is started when a
certain totalized flow is reached.
External start
The sampling program is started by a pulse
at the configured binary input.
External duration
The sampling program is active as long as
the configured input has the corresponding
level
Fieldbus (optional)
The sampling program is started by a signal
from the control system.
Having selected
Start condition
Date/time
Start date
01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time
00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Having selected
Start condition
Volume