Liquistation CSF34
Commissioning
Hauser
57
9.3.5
Sampling programs
Difference between program types
The following box provides an overview of the differences between the Basic, Standard and
Advanced program types.
Basic (1 sampling program)
Start condition:
• Immediate
• Date/time
• Immediate activation
• Time-paced, volume-paced or
flow-paced (CTCV, VTCV, CTVV),
external signal,
• Bottle change after time or
number of samples, external
signal
• Bottle synchronization
• Multiple bottles
Stop condition:
• Program end
• Continuous operation
Standard (1 sampling program with 1-5 sub-programs)
Start condition:
• Immediate
• Date/time
• Volume
• Immediate activation, individual
times, multiple times, interval,
deactivation of sub-program 1
• Time-paced, volume-paced or
flow-paced (CTCV, VTCV, CTVV),
external signal
• Bottle change after time or
number of samples, external
signal
• Bottle synchronization
• Multiple bottles
Stop condition:
• Program end
• Continuous operation
• Date/time
Advanced (1 sampling program with 1-24 sub-programs)
Start condition:
• Immediate
• Date/time
• Volume
• External signal
• Fieldbus
• Immediate activation, individual
times, multiple times, interval,
event, external start,
deactivation of sub-program 1,
fieldbus
• Time-paced, volume-paced or
flow-paced (CTCV, VTCV, CTVV),
single sample, sample table,
external signal, fieldbus
• Bottle change after time or
number of samples, external
signal, fieldbus
• Sample synchronization
• Bottle synchronization
• Multiple bottles
Stop condition:
• Program end
• Continuous operation
• Date/time
• Fieldbus