Liquistation CSF33
Troubleshooting
Hauser
31
347
Sampling
F
On
On
Sampling command not processed
1.
Check internal connection
2.
Reset the software
348
Read program
F
On
On
Selected program cannot be read from the pro-
gram memory
--> Create new program
349
Write program
F
On
On
Program created cannot be saved
Hardware error has occurred
--> Contact the Service Department
350
Module connect.
F
On
On
Cannot communicate with internal module
FMSY1
--> Check internal connecting cable
351
Delete program
F
On
On
Selected program cannot be deleted from the
program memory
--> Reset the software
352
Read prog. list
F
On
On
Program list cannot be read from the program
memory
--> Reset the software
353
Overfill check
M
On
Off
Total capacity of bottle reached
• No further sampling to current bottle is
triggered
354
Bottle check
M
On
Off
No empty bottles available for current program
• No further sampling
355
Start time over
M
On
Off
Start time entered is in the past
• Enter a new start time
356
Overfill check
S
On
Off
Volume of sample does not fit in the sample
bottle
--> Change the sample volume
357
No sampling
M
On
Off
Too many events at the current sampling time;
up to 24 sampling events can be triggered
simultaneously
--> Change the program settings so that samples
are taken at different times
358
Configuration
F
On
On
Program configuration does not match the
current device configuration
--> Adjust the configuration
359
Winter operation
M
On
Off
• Winter operation active
• Outside temperature too low
• No sampling
360
No sample
F
On
On
Intake hose blocked
--> Clean or replace the hose
361
Analog input
F
On
On
Current at analog input S:1 or S:2 too low
1.
Check current loop for error current
2.
Contact the Service Department
No.
Message
Factory settings
Tests or remedial measures
Cat.
Diag.
on/off
Error
current