Maintenance
Liquistation CSF33
10
Hauser
Interior of peristaltic pump
CAUTION
!
Danger of injury due to rotating parts
►
Take the sampler out of service before opening the peristaltic pump.
►
Safeguard the sampler against accidental operation while working on the open peristaltic pump.
1.
Take the sampler out of service by pausing a program that is currently running.
2.
Open the peristaltic pump as described in the "Replacing the pump tube" section.
3.
Remove the pump tube.
4.
Remove any silicone deposits on the roller and the flexible pump bracket.
5.
Make sure the roller turns smoothly and evenly.