Diagnostics menu
Liquistation CSF33
20
Hauser
2.4
Status of inputs/outputs
The following measured values are listed (read only):
• Binary inputs
Current function state: on or off
• Current inputs
Actual current values of all the current inputs available
• Binary outputs
Current function state: on or off
• Temperature sensors
Current value is displayed: S:1 (cooling system)
2.5
Systemtest/Reset
Diagnostics/Systemtest/Reset
Function
Options
Info
Power supply
Read only
The current supply voltage is displayed.
Cooling system (only for version with sample chamber temperature regulation)
Check cooling
Power supply
Read only
The current supply voltage is displayed.
24 V ±0.5 V
Overcurrent
Read only
No: no error
Yes: the fan in the climate control module is defective
-> Contact the Service Department
Temp. sample room
Read only
The
current
temperature of the sample compartment is
displayed.
Temp. sample room
Read only
When you start the cooling test, the temperature at the
start time is displayed
Cooling test off or Cooling test on -> progress of test is displayed
Start Test
Action
Start the cooling test.
Stop test
Action
End the cooling test.
Check heating
Power supply
Read only
The current supply voltage is displayed.
24 V ±0.5 V
Overcurrent
Read only
No: no error
Yes: the heating is defective
-> Contact the Service Department
Overcurrent
Read only
No: no error
Yes: the fan is defective
-> Contact the Service Department